85
________________________________________________________________________________________________________________
6.15 Commands Specific to the 3511-50
________________________________________________________________________________________________________________
:CORRection:DATA?
Queries the open circuit and short circuit compensation values.
:CORRection:OPEN
Enables and disables the open circuit compensation function.
Syntax
:CORRection:DATA?
<data>
Residual impedance: Numerical value in NR3 format or OFF (character data)
Phase angle: Numerical value in NR2 format or OFF (character data)
Function
Returns the open and short circuits compensation values in the current test frequency in
the following order.
<Residual impedance of short circuit compensation>
<Phase angle of short circuit compensation>
<Residual impedance of open circuit compensation>
<Phase angle of open circuit compensation>
When the compensation setting is OFF, returns the character data "OFF."
Example
Response
If headers are on
:CORRECTION:DATA OFF,OFF,247.45E+06,-21.58
If headers are off
OFF,OFF,247.45E+06,-21.58
The short circuit compensation for the current test frequency is set to OFF, and
open circuit compensation is 247.45 M
Ω
, -21.58
.
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
:CORRection:OPEN
<data>
<data>
ON/OFF (character data)
Function
Enables and disables the open circuit compensation function.
Specifying "ON" for <data> starts the reading of open circuit compensation data. Upon
completion of data reading, the open circuit compensation function is set to ON.
Example
Transmission
:CORRection:OPEN ON
The open circuit compensation function is set to ON.
Error
If <data> is other than character data described above, an execution error occurs.
Executing this command while the comparator function is performed generates an
execution error.
Summary of Contents for 3511-50
Page 1: ...INSTRUCTION MANUAL For 3511 50 LCR HiTESTER...
Page 2: ......
Page 137: ......
Page 138: ......