24
________________________________________________________________________________________________________________
3.8 Short Circuit Compensation
________________________________________________________________________________________________________________
NOTE
NOTE
NOTE
(2) Short circuit together the HIGH and LOW leads. In order to keep
external influences as low as possible, be sure to thrust the shorting bar
in all the way.
As closely as possible, route the test cable during this procedure as if
performing actual measurement.
(3) Press
. Begin compensation upon confirmation that the first
parameter measurement display (MAIN PARAMETER) is as shown.
The LED lamp flashes during short circuit compensation.
(4) If compensation ends properly, the unit beeps once and returns to normal
test mode.
(5) Refer to the LED lamp to the left of the key to check for normal
completion of short circuit compensation.
Short circuit compensation takes about 30 s.
If an error occurs during short circuit compensation, the unit beeps to notify
you of the error. An error message appears, as shown, and compensation is
aborted.
Press
to return the unit to normal test mode.
If an error occurs, when you return to normal test mode after aborting short
circuit compensation, the measurement value is not compensated for.