Test parameters
Impedance (Z), phase angle (
θ
)
Test frequency
1 kHz
Test signal level
1 V
Test range
AUTO
Equivalent circuit mode
AUTO
Open circuit compensation
OFF
Short circuit compensation
OFF
Trigger setting
Internal trigger
Test speed setting
NORMAL
Beep sound setting
ON for key input, OFF for comparator
Comparator
Both first and second parameters
Upper and lower limit values: OFF
Panel save
All contents clear
*
RST
Performs device initial setting.
*
TRG
Issues external trigger.
Syntax
*
RST
Function
Resets the 3511-50. The items which are reset are listed below.
Error
If the data parameters are set after this command, a command error occurs.
Syntax
*
TRG
Function
In external trigger mode, performs measurement once.
Example
Transmission :
TRIGger EXTernal;
*
TRG;:MEASure?
Error
Executing this command in internal trigger mode generates an execution error.
If the data parameters are set after this command, a command error occurs.
Executing this command while the open or short circuit compensation is performed
generates an execution error.