17
________________________________________________________________________________________________________________
3.6 Setting the Ranging
________________________________________________________________________________________________________________
1
2
3
4
5
6
7
8
9
10
11
12
13
14
A
Z-display range *
C-display range
Frequency
Range number
Common to
120 Hz and 1 kHz
Frequency
Range number
120 Hz
1 kHz
10
009.00 to 200.00 M
Ω
1
000.00 to 999.99 pF
00.000 to 99.999 pF
9
0.9000 to 9.9999 M
Ω
2
0.0000 to 9.9999 nF
000.00 to 999.99 pF
8
090.00 to 999.99 k
Ω
3
00.000 to 99.999 nF
0.0000 to 9.9999 nF
7
09.000 to 99.999 k
Ω
4
000.00 to 999.99 nF
00.000 to 99.999 nF
6
0.9000 to 9.9999 k
Ω
5
0.0000 to 9.9999
µ
F
000.00 to 999.99 nF
5
090.00 to 999.99
Ω
6
00.000 to 99.999
µ
F
0.0000 to 9.9999
µ
F
4
09.000 to 99.999
Ω
7
000.00 to 999.99
µ
F
00.000 to 99.999
µ
F
3
0.9000 to 9.9999
Ω
8
0.0000 to 9.9999 mF
000.00 to 999.99
µ
F
2
0.0900 to 0.9999
Ω
9
00.000 to 99.999 mF
0.0000 to 9.9999 mF
1
0.0100 to 0.0999
Ω
10
000.00 to 999.99 mF
00.000 to 99.999 mF
*:
Range of impedance which can be measured within the accuracy guaranteed
NOTE
If the impedance exceeds the measurement range, the first parameter display
is "UF" (underflow, i.e., impedance below the measurement range) or "OF"
(overflow, i.e., impedance above the measurement range), and the second
parameter display disappears.
Summary of Contents for 3511-50
Page 1: ...INSTRUCTION MANUAL For 3511 50 LCR HiTESTER...
Page 2: ......
Page 137: ......