Preventive Maintenance
Chapter 6: General Procedures
Function
Procedure
Leakage current
Unplug the power connector from the power outlet.
Connect the power connector of the bed to the test device and
connect this device to the power outlet.
Make sure that the bed is correctly supplied by the test device:
control unit LEDs lit (unlock the functions if necessary).
Take leakage current measures without operating any function.
The value must be less than 100 µA.
Examine the AC power cable and power supply unit if the value is
outside of the specifications.
Replace the AC power cable or power supply unit if necessary.
Patient pendant
Disconnect the patient pendant and examine the condition of the con
nector. Then reconnect or replace the pendant.
Press each of the controls to make sure that they activate the correct
function and they do not work intermittently. Each movement must
be continuous.
Replace the pendant if necessary.
Troubleshoot in the event of a doubt.
Hilow motors
Inspect the motors for the presence and tightness of the attachment
hardware. Replace as necessary.
Fully raise and lower the bed.
Make sure that there is no friction or abnormal noises and that no
audible overload indication can be heard during the movement.
Make sure that the Bed Not Down LED lights up on the control
pendant and goes out when the bed is in low position.
Replace the defective motor(s) in the event of a malfunction.
Troubleshoot in the event of a doubt.
Head section motor Examine the motor for the presence and tightness of the attachment
hardware. Replace as necessary.
Fully raise and lower the head section. Make sure that there is no
friction or abnormal noises and that no audible overload indication
can be heard during the movement.
Replace in the event of a malfunction.
Page 6 - 6
VersaCare™ Bed Service Manual (MAN333 REV 2)
Summary of Contents for VersaCare P3200
Page 2: ......
Page 6: ...Revisions NOTES Page iv VersaCare Bed Service Manual MAN333 REV 2 ...
Page 18: ...NOTES Page xvi VersaCare Bed Service Manual MAN333 REV 2 ...
Page 270: ...Exchange Policy Chapter 5 Parts List NOTES Page 5 4 VersaCare Bed Service Manual MAN333 REV 2 ...
Page 272: ...Warranty Chapter 5 Parts List NOTES Page 5 6 VersaCare Bed Service Manual MAN333 REV 2 ...
Page 279: ...Base Frame Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 13 ...
Page 281: ...Casters Chapter 5 Parts List NOTES 5 VersaCare Bed Service Manual MAN333 REV 2 Page 5 15 ...
Page 340: ...CPR Cable Chapter 5 Parts List NOTES Page 5 74 VersaCare Bed Service Manual MAN333 REV 2 ...
Page 363: ......
Page 365: ...Electronics Schematic P3200 Sheet 1 of 3 Chapter 3 ...
Page 366: ...Electronics Schematic P3200 Sheet 2 of 3 Chapter 3 ...
Page 367: ...Electronics Schematic P3200 Sheet 3 of 3 Chapter 3 ...
Page 368: ...Motor Control P C Board Schematic PN 68717 Chapter 3 ...
Page 369: ...Logic Control P C Board Schematic PN 69378 Chapter 3 ...
Page 370: ...Siderail Controls Schematic Chapter 3 ...
Page 371: ...Optional Controls Schematic Chapter 3 ...
Page 372: ...Patient Controls Schematic Chapter 3 ...
Page 373: ...Patient Pendant Schematic Chapter 3 ...
Page 374: ...Power Supply PN 72289 Sheet 1 of 2 Chapter 3 ...
Page 375: ...Power Supply PN 72289 Sheet 2 of 2 Chapter 3 ...
Page 376: ...Motor Control P C Board PN 72273 Chapter 3 ...
Page 377: ...Power Supply P C Board PN 70786 Sheet 1 of 2 Chapter 3 ...
Page 378: ...Power Supply P C Board PN 70786 Sheet 2 of 2 Chapter 3 ...