Information for OEM and Setup users | Commissioning
II
Extended TXT schema
The extended TXT file contains the deviations of the supporting points from the
nominal values on the X and Y axes.
The following example shows a grid with 5 x 5 supporting points with a spacing of
25 mm on the X axis and 20 mm on the Y axis.
Example
NLEC Data File
0.91
// Serial Number = CA-1288-6631-1710
MM
ON
Number of Grid Points (x, y):
5
5
Grid Block Size (x, y):
25.0
20.0
Offset:
0
0
Station (1, 1)
0.00000
0.00000
Station (2, 1)
0.00120
-0.00100
Station (3, 1)
0.00030
-0.00060
Station (4, 1)
0.00100
0.00160
Station (5, 1)
0.00210
0.00000
Station (1, 2)
0.00005
0.00200
Station (2, 2)
0.00130
0.00210
Station (3, 2)
0.00130
0.00220
Station (4, 2)
0.00050
0.00230
Station (5, 2)
-0.00040
0.00030
Station (1, 3)
-0.00010
-0.00020
Station (2, 3)
-0.00190
-0.00210
Station (3, 3)
-0.00010
0.00010
Station (4, 3)
0.00040
0.00210
Station (5, 3)
0.00190
0.00080
Station (1, 4)
108
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 06/2022
Summary of Contents for GAGE-CHEK 2000
Page 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 06 2022...
Page 3: ...I General information...
Page 7: ...1 Fundamentals...
Page 14: ...2 Safety...
Page 19: ...3 Basic operation...
Page 43: ...II Information for OEM and Setup users...
Page 48: ...1 Transport and storage...
Page 54: ...2 Mounting...
Page 60: ...3 Installation...
Page 72: ...4 Commissioning...
Page 121: ...5 Setup...
Page 150: ...6 File management...
Page 157: ...7 Settings...
Page 172: ...8 Servicing and maintenance...
Page 184: ...9 Removal and disposal...
Page 186: ...10 Specifications...
Page 194: ......
Page 195: ...III Information for the Operator user...
Page 240: ...14 What to do if...