Software Configuration
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 118
4-2 Just do a Visual Check
1.
If you want to execute a visual check (ignoring if it is due or not), go to
Start
>
Programs
>
MediCal Pro
>
MediCal Q/A Visual Check
.
2.
Follow each step on the screen until the application closes.
4-3 Run an individual task
1.
To run an individual task, start the MediCal Pro application by selecting
Start
>
Programs
>
MediCal Pro
>
MediCal Pro. 2.01
.
2.
Select “Q/A Setup” to have an overview of the available Q/A Tasks. Select the task you would
like to run by clicking on it. Right-click on the task and select “
Run Now
” in the popup menu to
execute that particular task. Follow further instructions until the execution of the task ends.
3.
The result of the execution is displayed in the “Last Result” column of the
Q/A setup view.
4-4 Look up histories of executed tasks
1.
If you want to view the history of a task, select
Start
>
Programs
>
MediCal Pro
>
MediCal
Pro. 2.01
).
2.
Select “
Q/A Setup
”. You now see the available tasks. Select the task for which you would like
to see the history. Right-click on it; a popup menu appears. Select “
History
…” to bring up the
history of results for that task.
Summary of Contents for Centricity RA600
Page 16: ...Introduction Page 16 Illustration 1 2 Computer Unit Front View ...
Page 24: ...Introduction Page 24 Modality Type Description IP Address S M GW Port AE Title ...
Page 206: ...Software Procedures SW 006 Single Media Archive SMA Install Configure Test Page 206 ...
Page 274: ...Renewal Parts Page 274 Jorge Aguirre med ge com Maurice Robert med ge com ...