Software Configuration
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 127
1.
In the Q/A section, the visual check task will state its last result is “NOT OK”.
2.
Go to the history of the Visual Check task. (See the section “Look up histories of executed
tasks” if you need help on this).
3.
You should see an overview of the results for the Visual Check task, where the last result (the
top one) is “NOT OK”. Select this result and click “
Details
” to get detailed information.
4.
The last result of the visual check task is visualized according to the displays in your display
system. If one or more displays had a negative result for the visual check task, you should see