Software Configuration
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 128
it here. Select the display (you can only select one display at a time) where the result is “NOT
OK” and click “
Details
”.
5.
The different questions of the visual check task are displayed with the corresponding answer
next to it. Here you will see what exactly caused the ‘NOT OK’ result of your visual check.
a.
If ‘NOT OK’ is next to the question “Verify that the 5% block can be seen in the center of the
0% block and that the 95% block can be seen in the center of the 100% block?”, then the
SMPTE test pattern was not displayed correctly. See the corresponding item for
troubleshooting.
b.
If ‘NOT OK’ is next to the question “Is the linearity acceptable?”, then there is a problem with
the linearity. See the corresponding item for troubleshooting.
c.
If ‘NOT OK’ is next to the question “Verify that there is a black pixel visible in the center of
each #?”, then there is a problem with the focus. See the corresponding item for
troubleshooting.