Troubleshooting
TSG 014 Workstation Application Problem Solving
Page 253
4. My HPs (hanging protocols)
aren’t listed in QC (quality
control)
Hanging protocols must be established in both the
Viewing
and
Quality
Control
tabs: Select
Viewing>Manage Automated Hanging Protocol...
or
Quality Control>Manage Automatic Hanging Protocol...
HP can be exported to disk from
Viewing
and imported into
QC
(or vice
versa). Detailed steps for export from
Viewing
follow:
1. Select
Viewing>Manage Automated Hanging Protocols...
2. In the
Name
window of the
Automated Hanging Protocols Manager
(Viewing)
window, highlight the desired hanging protocol name and click
Export
.
3. In the
Export Hanging Protocol to XML file
window, enter a name for
the selected hanging protocol and click
Browse
.
4. In the popup
Save As
window, navigate to the location where you want to
save the selected hanging protocol and click
Save
.
5. Click the
Export...
button to close the
Export Hanging Protocol to XML
file
window and click
OK
to close the
Automated Hanging Protocols
Manager (Viewing)
window.
6. Check the location of the new XML file.
Detailed steps for import to
Quality Control
follow
:
1. Select
Quality Control>Manage Automated Hanging Protocols...
2. In the
Import Hanging Protocol XML file
window, first check the Use
Automated Hanging Protocol check box, then click on the
Import...
button.
3. In the
Import Hanging Protocol XML file
window, click on
Browse
to find
the XML file you want to import.
4. Navigate to the file location and highlight the XML file and then click
Open
.
5. If desired, enter a new hanging protocol name for the incoming XML file,
then click the I
mport...
button. Note that the
Import succeeded
message
appears.
6. Check that the imported hanging protocol name appears in the list in the
Automated Hanging Protocols Manager
(QC) dialog box.
5. Centricity RA600 won’t allow
direct view
Check box must be selected for direct view.
Select
Configuration>Viewing>General
and check the
Direct View
box.
6. Cannot connect to film digitizer
Ensure that you are licensed for the acquisition module and that the digitizer
software has been installed and configured.
Problem
Possible Solution
Summary of Contents for Centricity RA600
Page 16: ...Introduction Page 16 Illustration 1 2 Computer Unit Front View ...
Page 24: ...Introduction Page 24 Modality Type Description IP Address S M GW Port AE Title ...
Page 206: ...Software Procedures SW 006 Single Media Archive SMA Install Configure Test Page 206 ...
Page 274: ...Renewal Parts Page 274 Jorge Aguirre med ge com Maurice Robert med ge com ...