A - 46 |
Airvo 3
- USER MANUAL
English
Pulse oximetry
Specifications are tabulated for the Airvo 3 and all compatible sensors unless otherwise stated.
Nonin:
Data update period
<30 sec
Measurement wavelengths and Output Power*
Red: 660 nanometers @ 0.8 mW max. avg.
Infrared: 910 nanometers @ 1.2 mW max avg. (using Nonin Purelight® sensor)
SpO
2
Accuracy (A
rms
**
)
No Motion
70 to 100%
Adults/Pediatrics***
Neonates
Reusable
8000AX Series:
800XJ Series:
8000SX Series:
8000R:
8000Q2:
± 2 digits
± 3 digits
± 2 digits
± 3 digits
± 3 digits
N/A
N/A
N/A
N/A
N/A
Disposable
6000CX Series:
7000X Series:
± 2 digits
± 2 digits
± 3 digits
± 3 digits
Motion
Reusable
8000AX Series:
800XJ Series:
8000SX Series:
Low Perfusion****
± 2 digits
± 3 digits
± 3 digits
± 2 digits
N/A
N/A
N/A
± 3 digits
Pulse Rate Accuracy
Adults/Pediatrics***
Neonates
No Motion (18–300 BPM)
Reusable
8000AX Series:
800XJ Series:
8000SX Series:
8000R:
8000Q2:
± 3 digits
± 3 digits
± 3 digits
± 3 digits
± 3 digits
N/A
N/A
N/A
N/A
N/A
Disposable
6000CX Series:
7000X Series:
± 3 digits
± 3 digits
± 3 digits
± 3 digits
Motion (40–240 BPM)
Reusable
8000AX Series:
800XJ Series:
8000SX Series:
Low Perfusion (40–240 BPM)****
± 5 digits
± 5 digits
± 5 digits
± 3 digits
N/A
N/A
N/A
± 3 digits
* This information is especially useful for clinicians performing photodynamic therapy.
**
± 1 A
rms
represents approximately 68% of measurements.
*** Includes Infant patients
**** Does not apply to those sensors listed as N/A under the neonate column, 8000R and 8000Q2
Notes:
• SpO
2
accuracy testing is conducted during induced hypoxia studies on healthy, non-smoking, light-to-dark-skinned subjects during motion and
no-motion conditions in an independent research laboratory. The measured arterial hemoglobin saturation value (SpO
2
) of the sensors is
compared to arterial hemoglobin oxygen (SaO
2
) value, determined from blood samples with a laboratory co-oximeter. The accuracy of the
sensors in comparison to the co-oximeter samples measured over the SpO
2
range of 70–100%. Accuracy data is calculated using the root-mean-
squared (Arms value) for all subjects, per ISO 80601-2-61 formerly ISO 9919, Standard Specification for Pulse Oximeters for Accuracy.
• Pulse rate motion testing measures pulse rate accuracy with motion artifact simulation introduced by a pulse oximeter tester. This test
determines whether the oximeter meets the criteria of ISO 80601-2-61, formerly ISO 9919, for pulse rate during simulated movement, tremor, and
spike motions.
• Low perfusion testing uses an SpO
2
Simulator to provide a simulated pulse rate, with adjustable amplitude settings at various SpO
2
levels. The
module must maintain accuracy in accordance with ISO 80601-2-61, formerly ISO 9919, pulse rate and SpO
2
at the lowest obtainable pulse
amplitude (0.3% modulation).
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......