5.4 During therapy
Monitor the patient following hospital protocols and clinical judgement. Ensure you can hear and respond to any device alarms.
If there is an interruption to the power supply, and the battery is depleted, the Airvo 3 will raise a Power Out alarm, turn off , and not
deliver any therapy to the patient. The power out alarm will sound once every 10 seconds for a minimum of 120 seconds, and the signal
light above the touch screen will fl ash. Once power is restored the Airvo 3 can be restarted and will retain the previous therapy and alarm
settings.
Warning
If using the battery as the power source, periodically check the battery status to ensure the battery does not become depleted
while therapy is being delivered.
5.4.1 Lock Screen (optional)
The lock screen can prevent accidental settings changes.
To enable the lock screen:
1. Tap to open the system menu.
2. Select Lock screen from the system menu.
The symbol is shown in the Message bar.
To disable the lock screen:
3. Touch the screen while it is locked, then hold the Unlock icon for
three seconds.
Stabilizing to target
value
Target
value
Unit
5.4.2 Monitor and adjust settings
Adjust settings as needed. Most changes take eff ect after pressing the confi rmation
button but it may take a few minutes for some settings, such as target humidity, to
respond to changes. Tiles show an animated ellipsis symbol (…) to indicate that a therapy
setting has not yet reached its target.
5.4.3 Manage condensation
Drain excess condensate from the breathing tube by:
1. Disconnecting the breathing tube from the patient interface, and
2. Lifting the patient end of the tube so the condensate runs into the water chamber.
Reduce the fl ow rate below 30 L/min if the condensate does not run freely into the water chamber. Return the fl ow rate to the prescribed
setting after draining the breathing tube.
Direct cold air away from the heated breathing tube where possible. Air conditioners, fans, open windows and other sources of cold air
may increase condensation.
Consider reducing the target humidity if condensation persists.
1
2
3
3 sec
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......