4.3.2 Low-pressure oxygen (LPO) inlet port
When using the LPO port, the amount of oxygen taken in by the Airvo 3 is controlled by an external
fl owmeter. Connect a tube from the external fl owmeter to the LPO port. Make sure that the fl owmeter
is turned off whenever the Airvo 3 is not delivering therapy.
When using the low-pressure oxygen inlet port, monitor the oxygen concentration displayed on the
Home screen. The oxygen fl ow-regulator must be adjusted manually to maintain the prescribed
oxygen concentration when changing the respiratory gas fl ow rate.
Clinicians may confi gure a High FiO
2
alarm to discourage use of high FiO
2
values in particular clinical
environments.
The High FiO
2
alarm can be disabled or a threshold between 30% and 95% can be selected when the
Airvo 3 is initially set up for your environment (see Oxygen high alarm threshold, Airvo 3 Technical
Manual). The alarm threshold is displayed on the Titrate FiO
2
screen, if enabled. Tap the FiO
2
tile to
open the Titrate FiO
2
screen.
5. Using the Airvo 3
5.1 Getting started
Turn on the Airvo 3
Plug the Airvo 3 power cord into the wall power supply.
Lock the wheels of the mobile pole stand to prevent the Airvo 3 from
moving.
Turn on the Airvo 3 by holding down the Power on/off button for
2 seconds.
Note
If the Airvo 3 has been unused and disconnected from the wall power
supply for some time, the device will not power on without being
plugged in.
Wall oxygen
supply
Flowmeter
Warning
Turn off the low-pressure oxygen source whenever the Airvo 3 is not delivering therapy, to ensure that oxygen does not build up
inside the device.
Warning
Make sure the Airvo 3 is dry before plugging the power cord into the
wall power supply to avoid a potential electric shock.
Warnings
The Airvo 3 must be cleaned and disinfected between patients. Refer to section 8 for the steps required to reprocess the Airvo 3
between patients.
Do not exceed the maximum use period for single-patient-use accessories and consumables (see section 8.3 for the schedule for
changing accessories).
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......