8. Reprocessing
Standard aseptic techniques should be followed to minimize contamination when handling the Airvo 3 and accessories. The patient
interface, heated breathing tube, water chamber and outlet elbow may become contaminated during use. As soon as possible after using
the Airvo 3:
1. Remove the single-use accessories from the Airvo 3 and dispose of them in accordance with local laws, regulations and hospital
protocols for disposing of contaminated products.
– Squeeze the sides of the breathing tube connector and lift to remove it from the Airvo 3.
– Grip the port adapter and pull the water chamber away from the Airvo 3 to remove it.
2. Reprocess the Airvo 3 device exterior by following the instructions in section 8.1.
3. Clean and high-level disinfect the Outlet Elbow by following the instructions in section 8.2.
4. Replace accessories within the maximum use period shown in section 8.3 (schedule for changing accessories).
5. Clean and disinfect pulse oximetry accessories (including reusable sensors) in accordance with the manufacturer’s instructions.
8.1 Airvo 3 device exterior reprocessing
8.1.1 Clean
Equipment
• Mild detergent
• Clean, disposable, lint-free cloths
• Protective gloves
Cleaning Instructions
1. Mix a solution of warm water and mild detergent (refer to the detergent manufacturer’s instructions for use).
2. Dampen a clean cloth with the cleaning solution.
3. Thoroughly wipe all outside surfaces of the device (including the Outlet Elbow) for at least one minute to remove any visible soil. Use
the corner/edge of the cloth to clean all crevices of the device.
Rinse
4. Dampen a clean cloth with tap water.
5. Thoroughly wipe all outside surfaces of the device with the damp cloth to remove any disinfectant residue.
Dry
6. Thoroughly wipe all outside surfaces of the device with a dry cloth until it is visibly dry.
7. Allow to air dry until completely dry.
Warnings
Do not clean and/or disinfect the Airvo 3 while it is being used by a patient.
Do not submerge the Airvo 3 or accessories in any cleaning solution or attempt to sterilize by autoclave, irradiation, steam, gas,
ethylene oxide or any other method. This will seriously damage the device.
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......