Airvo 3
- USER MANUAL | A - 27
English
7. Troubleshooting
This section describes common causes, and solutions to, problems and alarms that you may encounter while using the Airvo 3. The
Airvo 3 Technical Manual contains additional information that may be helpful in resolving more advanced problems.
7.1 Alarms
The Airvo 3 has visual and auditory alarms to notify users about interruptions to a patient’s treatment. These alarms are generated by an
intelligent alarm system, which processes information from sensors and target settings of the device and compares this information with
pre-programmed limits. Changes to alarm settings will be preserved during or after any power loss.
The signal light fl ashes and troubleshooting information is displayed on the Airvo 3
touch screen when any alarm is active. The color of the signal light indicates the
highest-priority active alarm condition.
7.2 Alarm priority
Alarms are grouped by urgency and severity into three priority levels: low, medium, high. When multiple alarms are active, the audible
alert, signal light and Message bar background color will signal the highest-priority alarm active.
• A response is needed for all alarms.
• A prompt response is required for all medium-priority alarms.
• An immediate response is required for all high-priority alarms.
Priority
Message bar, signal light color
Audible alert
Low
Solid yellow
High then low-pitched beep
Medium
Flashing yellow
3 beeps every 9 seconds
High
Flashing red
10 beeps every 5 seconds
Warning
Audible alarms may not be heard if the alarm volume is set lower than ambient noise. Missed alarms may lead to patient injury. Refer
to the Airvo 3 Technical Manual to review and set the alarm volume.
7.3 Auditory information signals
The informative sounds made by the Airvo 3 are:
Melody
Meaning
Ascending sequence of 5 tones
The breathing gas has warmed up
Single tone
A touch on the display has been registered
Single low then high tone
All active alarms have been resolved
High note followed by 2 (identical) lower notes,
repeated every 10 seconds
The Power Out alarm is active. The wall power supply has been
disconnected or turned off and the (optional) battery is depleted
Descending sequence of 3 tones
The device has completed the power off process
Sequence of 3 tones with high, low then middle pitch
The device has been turned on
Signal light
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......