Airvo 3
- USER MANUAL | A - 29
English
Alarm condition
Priority
Delay
Meaning
Unsupported Battery
Med.
<5 s
The device is running off the battery and either an incorrect battery type is
connected or communications with the battery could not be established.
Charging is disabled. During battery use the behaviour is the same as the
Critically Low Battery alarm.
Critically Low Battery
Med.
<5 s
The Airvo 3 battery level is critically low. Connect the Airvo 3 to a wall power
supply immediately to maintain therapy. Humidifi cation is turned off to maintain
operation of the blower and oxygen supply.
Low Battery
Low
<5 s
The Airvo 3 battery level is low and should be connected to a wall power supply.
Therapy continues as normal.
Battery Mode:
Reduced Humidity
Low
<5 s
The Airvo 3 has been disconnected from the wall power supply and the
device is now running off the battery. The delivered humidity may be
reduced.
Battery Charger Fault
Low
<30 s
The battery charger is not functioning correctly and has been disabled. Restart
the device to resolve the fault. If the issue persists, contact your service
representative.
Therapy alarm - tube
Outlet Elbow Missing
High
<15 s
The Airvo 3 outlet elbow has been removed from the device during therapy.
Check that the outlet elbow is fully inserted into the Airvo 3. If the issue persists,
replace the outlet elbow.
Check Tube
Med.
<5 s
The Airvo 3 cannot detect the heated breathing tube. Check that the heated
breathing tube is not damaged and is plugged in correctly. Replace the heated
breathing tube if the problem persists.
Wrong Tube
Med.
<5 s
The heated breathing tube is not suitable for the selected therapy, or is
damaged. Connect a suitable heated breathing tube. Replace the breathing tube
if the problem persists.
Outlet Elbow Fault
Med.
<5 s
A fault has been detected with the outlet elbow. Check that the outlet elbow is
fully inserted into the Airvo 3. If the issue persists, replace the outlet elbow.
Outlet Elbow Too Warm
Med.
<5 s
The outlet elbow is too warm to run start up checks. Wait for the outlet elbow to
cool down. If the issue persists, replace the outlet elbow.
Therapy alarm – high fl ow
Chamber Leak Detected
Med.
<30 s
The water chamber has been removed. Ensure the water chamber is
correctly inserted into the Airvo 3. If the issue persists, contact your service
representative.
Leak Detected
Med.
<30 s
The Airvo 3 has detected a leak in the breathing circuit. Check:
• the water chamber has not been removed and is properly installed,
• the heated breathing tube is plugged in correctly or is not damaged,
• the patient interface is fi tted correctly, and
• the air fi lter is fi tted correctly.
If the issue persists, replace the consumables.
Blockage Detected
Med.
<15 s
The Airvo 3 has detected a blockage. Check:
• for blockages in the heated breathing tube, patient interface and inlet air
fi lter,
• the patient interface is the correct size for the patient, and
• the target fl ow rate is within the rated range of the interface.
If the issue persists, replace the consumables.
Flow Below Target
Med.
<2 min
The Airvo 3 fl ow rate is lower than the target fl ow rate. Check:
• for blockages in the heated breathing tube, patient interface and inlet air
fi lter,
• the patient interface is the correct size for the patient, and
• the target fl ow rate is within the rated range of the interface.
If the issue persists, replace the consumables.
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......