A - 32 |
Airvo 3
- USER MANUAL
English
Alarm condition
Priority
Delay
Meaning
Check Tube
Med.
<5 s
The Airvo 3 cannot detect the disinfection tube. Check that the disinfection
tube is not damaged and is plugged in correctly, then restart the device.
If the problem is not resolved, replace the disinfection tube and outlet elbow in
turn. If the issue persists, contact your service representative.
Leak Detected
Med.
<35 s
The Airvo 3 has detected a leak in the disinfection circuit. Check:
• The disinfection tube blue connector is connected to the top of the outlet
elbow,
• The disinfection tube red end is connected to the left hand chamber port,
• The disinfection filter is connected to the right hand chamber port,
Then restart the device. If the problem is not resolved, replace the disinfection
tube and outlet elbow in turn. If the issue persists, contact your service
representative.
Blockage Detected
Med
<10 s
The Airvo 3 has detected a blockage. Check that the disinfection tube is not
blocked and that the disinfection filter is not wet, then restart the device.
If the problem is not resolved, replace the disinfection tube and outlet elbow in
turn. If the issue persists, contact your service representative.
Check Operating
Conditions
Med.
<1 min
The Airvo 3 has detected ambient conditions that are not suitable. Do not use
the Airvo 3 when the ambient temperature is below 18 °C or above 28 °C.
Move the device to a suitable environment, then restart the device. If the issue
persists, contact your service representative.
Wall Power Disconnected
Med.
<5 s
The power cable has been removed and device is unable to perform a
disinfection cycle. Connect device to wall power, then restart the device. If the
issue persists, contact your service representative.
Unexpected O
2
Med.
<1 min
Oxygen is being supplied to the Airvo 3 while in disinfection mode. Check all
oxygen supplies are turned off and disconnected. If the issue persists, contact
your service representative.
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......