A - 38 |
Airvo 3
- USER MANUAL
English
Pulse oximetry measurements and status are shown as follows:
Tap the Pulse Oximetry tile to open the Pulse Oximetry screen.
SpO
2
plethysmograph
% SpO
2
Pulse rate
High/low alarm setting
Open Pulse Oximeter
Alarms and Settings
page
Warnings
Nonin:
The Nonin Xpod USB connector is designed to determine the percentage of arterial oxygen saturation of functional
haemoglobin. Factors that may degrade pulse oximetry performance or aff ect the accuracy of the measurement
include the following:
• Excessive ambient light
• Excessive motion
• Electrosurgical interference
• Elevated levels of dyshemoglobin
• Blood fl ow restrictors (arterial catheters, blood
pressure cuff s, infusion lines, etc.)
• Moisture in the sensor
• Improperly applied sensor
• Incorrect sensor type
• Poor pulse quality
• Venous pulsations
• Anemia or low haemoglobin concentrations
• Cardiogreen or other intravascular dyes
• Carboxyhemoglobin
• Methemoglobin
• Dysfunctional haemoglobin
• Artifi cial nails or fi ngernail polish
• A sensor not at heart level
Loss of monitoring can result if any objects hinder the pulse measurement. Ensure that no blood fl ow restrictors (e.g.,
blood pressure cuff ) hinder pulse measurements.
The oximeter sensor may not work on cold extremities due to reduced circulation. Warm or rub the fi nger to increase
circulation, or reposition the sensor.
Inaccurate readings can result due to residue (e.g. dried blood) in light path or degradation of optical characteristics of
sensor components. Refer to cleaning instructions supplied with the pulse oximetry accessories.
False high readings can result if SpO₂ is low due to dysfunctional hemoglobin (e.g. carboxyhemoglobin or
methemoglobin).
Read the instructions supplied with the pulse oximetry accessories for additional safety information (including any
potential risks or adverse eff ects from sensor materials), measurement site selection, detailed sensor setup, maximum
sensor application time at a single site before repositioning, cable lifetime, sensor lifetime, factors that can interfere
with measurement, troubleshooting and maintenance instructions.
Pulse oximetry is
connected to the Airvo 3,
but the sensor is not
attached to a patient.
Poor signal quality may
be causing inaccurate
measurements
(see section 7 for
troubleshooting).
High/low SpO
2
alarm threshold
(“--” if disabled)
Pulse
rate
SpO
2
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......