Airvo 3
- USER MANUAL | A - 31
English
Alarm condition
Priority
Delay
Meaning
High FiO
2
(HPO)
Low
<5 s
The FiO₂ target is above the selected High Oxygen Alarm threshold for its
intended clinical environment (range 30–95% or Off, default: 95%, see Airvo 3
Technical Manual). Check FiO
2
is appropriate for the patient’s condition. Reduce
FiO
2
to the normal range when it is appropriate to do so.
Pulse oximetry alarms
Pulse Ox Disconnected
Med.
<1 s
The pulse oximetry accessories have become disconnected. Reconnect pulse
oximetry accessories.
Pulse Ox Communication
Failure
Med.
<10 s
The Airvo 3 is unable to communicate with the pulse oximeter. Check that the
USB connector cable, sensor adapter cable and sensor cables are all properly
connected. Replace the sensor cable, adapter cable then USB connector cable in
turn if the problem persists.
No Pulse Ox Sensor
Connected
Med.
<1 s
A pulse oximetry sensor cable was not detected or is inoperable. Check that the
sensor cable is properly connected to the USB connector cable or replace the
sensor cable if necessary.
Pulse Ox Sensor Off
Patient
Med.
<1 s
The pulse oximeter is no longer receiving SpO
2
measurements from the patient.
Check that the sensor is properly attached to a suitable measurement site on
the patient.
No SpO
2
Reading
Med.
<16 s
The pulse oximeter is not sending valid SpO
2
measurements.
Check the sensor, cable and USB interface. Try replacing each component in turn
until the problem is resolved.
No Pulse Rate Reading
Med.
<16 s
The pulse oximeter is not sending valid Pulse Rate measurements.
Check the sensor, cable and USB interface. Try replacing each component in turn
until the problem is resolved.
Pulse Ox Not Recognized
Med.
<10 s
The selected pulse oximeter has not been recognised. Please remove or change
oximeter.
Physiological alarms
Low SpO
2
High
<15 s
Check your patient. The SpO
2
measurement has decreased below the Low SpO
2
alarm threshold. Check that the alarm setting is appropriate for your patient
(Range: 1–98%, default 85%, see Airvo 3 Technical Manual).
High SpO
2
Med.
<15 s
Check your patient. The SpO
2
measurement has exceeded the High SpO
2
alarm
threshold. Check that the alarm setting is appropriate for your patient (Range:
2–99% or Off, default Off, see Airvo 3 Technical Manual).
Disinfection alarms
Disinfection Failed to
Hold Temperature
Med.
<3 min
The Airvo 3 cannot heat up to the required disinfection temperature. Check:
• the disinfection tube blue connector is connected to the top of the outlet
elbow,
• the disinfection tube red end is connected to the left hand chamber port,
• the disinfection filter is connected to the right hand chamber port,
Then restart the device. If the problem is not resolved, replace the disinfection
tube and outlet elbow in turn. If the issue persists, contact your service
representative.
Over Temperature
Detected
Med
<5 s
The Airvo 3 detected higher than expected temperatures during the
disinfection cycle. Check:
• the disinfection tube blue connector is connected to the top of the outlet
elbow,
• the disinfection tube red end is connected to the left hand chamber port,
• the disinfection filter is connected to the right hand chamber port,
Then restart the device. If the problem is not resolved, replace the disinfection
tube and outlet elbow in turn. If the issue persists, contact your service
representative.
Summary of Contents for Airvo 3
Page 1: ...www fphcare com USER MANUAL...
Page 2: ...SECTION English A Espa ol B C Fran ais Canadien D...
Page 116: ...Airvo 3 C 7 3 Airvo 3 3 1 3 1 Airvo 3 AirSpiral Optiflow Airvo 3...
Page 117: ...C 8 Airvo 3 3 2 HPO LPO USB 1 USB 2 HPO DISS NIST SIS...
Page 118: ...Airvo 3 C 9 3 3 Airvo 3 3 3 1...
Page 119: ...C 10 Airvo 3 3 3 2 7 3 3 3 Airvo 3 50 50 Airvo 3 USB USB 3 3 4 7...
Page 126: ...Airvo 3 C 17 Airvo 3 1 2 1 Airvo 3 Airvo 3 2...
Page 130: ...Airvo 3 C 21 Airvo 3 EEG ECG EKG EMG Airvo 3...
Page 131: ...C 22 Airvo 3 5 4 Airvo 3 10 120 Airvo 3 5 4 1 1 2 3 5 4 2 5 4 3 1 2 30 L min 1 2 3 3...
Page 133: ...C 24 Airvo 3 5 6 1 2 Airvo 3 Airvo 3 1 2 3 2 Airvo 3 4 Airvo 3 Airvo 3 2 3 4 1 2...
Page 134: ...Airvo 3 C 25 6 Airvo 3 24 Airvo 3 Airvo 3 Airvo 3 F P Airvo 3 F P Airvo 3 Airvo 3 Airvo 3...
Page 136: ...Airvo 3 C 27 7 Airvo 3 Airvo 3 7 1 Airvo 3 Airvo 3 7 2 9 3 5 10 Airvo 3 7 3 Airvo 3 5 2 10 3 3...
Page 141: ...C 32 Airvo 3 5 Airvo 3 35 Airvo 3 10 Airvo 3 1 Airvo 3 18 C 28 C Airvo 3 5 1 Airvo 3...
Page 147: ...C 38 Airvo 3 SpO2 SpO2 Nonin Nonin Xpod USB Cardiogreen SpO Airvo 3 7 SpO2 SpO2...
Page 149: ...C 40 Airvo 3 9 3 2 PR bpm 9 3 3 9 3 4 Nonin Nonin Airvo 3 SpO2 9 4 9 5 9 4 1 SpO2 SpO2 SpO2...
Page 152: ...Airvo 3 C 43 9 6 SpO2 Airvo 3 Airvo 3 1 SpO2 2 SpO2 SpO2 SpO2...
Page 157: ...C 48 Airvo 3 USB USB II MR Type BF BF REF REF 93 42 EEC UL...
Page 161: ...C 52 Airvo 3 4 Airvo 3 Airvo 3 Optiflow Airvo 3 Airvo 3 Airvo 3 Airvo 3 Optiflow 12 mg L 40 20...
Page 215: ......