IEC STANDARD 60601-1-2:2014, 4th Edition
This device is approved for use in a specifi c electromagnetic environment. The customer or user of the
device must ensure that it is used in an electromagnetic environment in accordance with the description
given below.
Emission Measurement
Agreement
Guidelines Regarding the Electromagnetic Environ-
ment
RF-emission according to
CISPR 11
Group 1
This device uses RF-energy for internal functions only. RF-
emissions are therefore very low, and it is unlikely that other
nearby electronic equipment will be disturbed.
RF-emission according to
CISPR 11
Class B
The device is suitable for use in all environments, including
residential areas, and approved for direct connection to the
public low-voltage network for residential areas.
Harmonics according to
IEC 61000-3-2
Class A
Voltage fl uctuations/ fl icker
according to IEC 61000-3-3
met
Interference Immunity
Test
IEC 60601-
test level
Compliance level
Electromagnetic Environment - Guidelinesf
Electrostatic
Discharge (ESD) according
to IEC 61000-4-2
± 8 kV contact
discharge
±2, 4, 8, 15 kV air
discharge
± 8 kV contact discharge
±15 kV air discharge
The fl oor should be wood, concrete or tiles. If the
fl oor is covered with synthetic material, the relative
humidity should be at least 30%. Criterion B
Fast transient electrical dis-
turbances/bursts according
to IEC 61000-4-4
(only for V 300/600)
± 2 kV for mains
100 kHz repeat rate
± 2 kV for mains
100 kHz repeat rate
The quality of the mains power supply should
meet the requirements for a normal commercial or
clinical environment. Criterion B
Surge voltages according to
IEC 61000-4-5
(only for V 300/600)
± 0,5 kV , ± 1 kV
L to N
± 0,5 kV , ± 1 kV ± 2 kV
L to GND
± 1 kV
L to N
± 2 kV
L to GND
The quality of the mains power supply should
meet the requirements for a normal commercial or
clinical environment. Criterion B
Voltage dips, short-term
interruptions and voltage
fl uctuations of the mains
supply lines according to
IEC 61000-4-11
(only for V 300/600)
0 % UT
0°,45°,90°,135°,180°,2
25°,270°,315°
0 % UT
0°
0%
70 % UT
0 % UT
0%
0 % UT
for 1/2 Period
1 Period
25 /30
Periods
(50/60Hz)
250/300 Periods
(50/60Hz) for 5 s
The quality of the mains power supply should meet
the requirements for a normal commercial or clini-
cal environment. If the user of the product requires
continuous operation even with interruptions of the
power supply, the product should be connected to
an uninterruptible power supply.
Criterion A (max. mains voltage)
Criterion B (min. mains voltage)
Criterion A (max. mains voltage)
Criterion B (min. mains voltage)
Criterion A (max. mains voltage)
Criterion B (min. mains voltage)
Criterion A (max. mains voltage)
Criterion B (min. mains voltage)
Magnetic fi eld at the mains
frequency (50/60 Hz) accor-
ding to IEC 61000-4-8
30 A/m
30 A/m
Magnetic fi elds at the mains frequency should
have levels typical of an application in a
commercial or clinical environment.
Note: UT is the alternating mains voltage prior to the application of the test level.
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...