6.1 — CONNECTION TO ELECTRICAL SYSTEM
The connection involves hooking up the two electrical
terminals to the transformer, arranged to provide an output
of 24 V (see wiring diagram 14.1). Model L only: voltage for
LED light 3.3 Vdc: black wire (+), white wire (-)
NOTE
• The electric power supply and transformer used must
be rated Safe Extra Low Voltage (SELV) for medical use, in
compliance with the mandatory regulations.
• Connection to terminals with output voltage exceeding 24
V could irreparably damage the unit and compromise safety.
6.2 — CONNECTION TO HYDRAULIC SYSTEM
The syringe’s green line must be hooked up to the water
supply.
NOTE
• The operating pressure is 250 kPa. A slight increase in
pressure would only increase the power of the jet, but
would not create any problems and is not dangerous.
• The water used must be potable water, filtered (<25 μm)
and free of bacteria, etc.
• For the syringe to function properly, the water pressure
must not be lower than indicated.
6.2 — CONNECTION TO COMPRESSED AIR SYSTEM
The syringe’s blue line must be connected to the compressed
air system.
NOTE
• The recommended operating pressure is around 450
kPa.
• When using the Minilight syringe, national regulations
regarding water and air quality must also be met.
• The air must be medical grade, dry and free of oil and
bacteria – a 5μm air filter is recommended.
6.4 — CONNECTION OF STRESS ABSORBER CORD
The stress absorber cord must be anchored to the dental
5.1 — OPERATING CONDITIONS
5.2 — TRANSPORT AND STORAGE CONDITIONS
5 —— TECHNICAL CHARACTERISTICS
6 —— INSTALLATION AND
CONNECTIONS
FUNCTIONS
6F
3F
L
Supply voltage
V~
24
***
24
LED voltage
Vdc
***
***
3,5
Max. absorbed
current
A
4,3
***
4,3
Electrical power
W
103
***
103
Water supply
pressure
Kpa
250
250
250
Air supply
pressure
Kpa
450
450
450
Water flow rate
Cc/min
110
110
110
Air flow rate
NI/min
10
10
10
International
protection
IP40
IP40
IP40
Ambient temperature
10°C / +45°C
Relative air humidity
30% / 85%
Atmospheric pressure
80 Kpa - 106 Kpa
Temperature
-20°C / +60°C
Relative Humidity
30% / 85%
Atmospheric Pressure
50 Kpa - 106 Kpa
IMPORTANT NOTE:
With each syringe, the package also includes a User’s manual
which also includes a “Fault Report Form”. Since this is
required by law, the user must receive this User’s manual.
Therefore, the syringe installer is responsible for delivering
these forms to the dentist. Directive 93/42/EEC requires
product traceability: therefore, our customers are asked to
ensure that we can identify the end customer to whom the
product has been sold.
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...