EVA
USER MANUAL - EN
Edition 1.0 April 2020
Pag. 11 di 14
6
CLEANING AND DISINFECTION
Warning against danger of wear and corrosion and falling suspended mass
For all metal or plastic parts it is strictly forbidden to use substances that are
-
abrasive,
-
corrosive,
-
acids,
-
substances containing chlorine or chloride ions, phosphorous or phosphorous ions,
-
detergents with Trilene base, petrol, white spirit, chlorine or similar.
Do not use detergents-disinfectants containing the following substances to clean plastic parts:
-
Ammonium Hydroxide
-
Sodium Hydroxide
-
Hydrogen peroxide
-
Ammonium Chloride
-
Methylene chloride
-
Methyl alcohol
-
Acids and corrosive substances of all kinds.
It is forbidden to directly spray any chemical substance on the device.
It is forbidden the use of wet wipes without rinsing.
Faro tested and suggests the use of the following disinfectants, for plastic parts and metal parts:
-
Durr FD366 Sensitive
-
Faro Perflex Advanced
-
Water-alcohol based disinfectants with 70% isopropyl alcohol or ethanol are suitable.
6.1
CLEANING OF THE REFLECTING PARABOLAS
Cleaning must be carried out using a soft cloth in cotton or absorbent cotton with ethyl alcohol.
Water-alcohol based disinfectants are suitable with 70% isopropyl alcohol or ethanol.
Caution - potential damage or wear on the parabolas
Never spray detergent directly on the parabolas.
Cleaning operations on the parabolas must be carried out wearing gloves, to avoid leaving fingerprints on the surfaces.
Never use detergents containing surfactants or water-repellents that depositing can leave streaks. Slight streaking will not
prejudice the quality of the light.
Products differing from those suggested could damage the parabolas.
If in doubt, contact FARO customer care.
6.2
CLEANING AND DISINFECTION OF THE HEAD
Cleaning must be carried out using a soft cloth in cotton wetted with disinfectant solution.
Always squeeze the cloth to remove all the liquid in excess.
6.3
CLEANING AND DISINFECTION OF ARMS
Always use a cloth soaked in disinfectant approved to disinfect the surfaces and pass it over.
Always squeeze the cloth to remove all the liquid in excess.
7
STERILIZATION OF THE HANDLES
Warning - danger of cross contamination
The handles are not supplied sterile, they must therefore be sterilised before
use. The handles must be sterilised before each patient.
7.1
REMOVAL OF THE HANDLES
To remove the handle, unscrew knob “A” and remove it from the support.
7.2
DECONTAMINATION AND DISINFECTION
Before sterilising the handles, they must be decontaminated and disinfected.
To disinfect, Faro has tested the following products for disinfection:
Faro Perflex Advance
Durr FD366 Sensitive
WARNING - danger of plastic breaking
The handles cannot be disinfected by thermo-disinfection.
7.3
STERILIZATION
The handles must be packaged in compliance with EN 868-5.
The handles can be sterilised with standard cycles 121°/134° C up to two hundred (200) cycles or however up to loss of the
mechanical performance.
The parameters of the sterilisation cycle are as follows:
Cycle EN 13060
Temperature
Pressure
Holding Time Minimum
B
121°C
207 kPa
15 min
B
134°C
308 kPa
3 min
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...