26
7117100018L30 1712V002
Troubleshooting
Fault
Probable cause
Solution
Yellow display is on
GREEN LED illuminates
Audible signal melody
sounds
Amalgam collecting container is
95% full
i
Change the amalgam collecting
container�
Float sensor dirty or blocked
i
If this display occurs repeatedly
even when the collecting con-
tainer is empty, check that the
float sensor can move freely�
Yellow display is on
Red display flashes
Audible signal melody
sounds
Amalgam collecting container is
100% full
i
Change the amalgam collecting
container� Audible signal can no
longer be switched off�
Float sensor dirty or blocked
i
If this display occurs repeatedly
even when the collecting con-
tainer is empty, check that the
float sensor can move freely�
Waster water line/siphon trap
dirty
i
Clean the waste water line/si-
phon trap� *
The RED and GREEN dis-
plays flash alternately
Audible signal
Motor is dirty or defective
i
Check motor for smooth run-
ning; replace the centrifuge if
necessary� *
i
Replace the device� *
Contact problems at X9
i
Plug in the connector correctly� *
i
Replace the PCB main board
and connector on the motor� *
Orange LED flashes
Audible signal
Press the service key briefly to
switch off the audible signal
Amalgam collecting container not
correctly in position
i
Switch OFF the device�
i
Insert the amalgam collecting
container in the correct position�
i
Switch ON the device�
Float sensor missing
i
Insert the float sensor� *
Water accumulating in the
spittoon
Coarse sieve in the fluid inlet
blocked
i
Clean the coarse sieve�
Outlet ineffective or not vented
i
Check or retrofit the ventilation� *
Suction power too weak or
interrupted
Coarse sieve is blocked on the
inlet of the aspiration
i
Clean the coarse sieve�
Place selection valve not or
incompletely open
i
Check the control voltage� *
i
Clean the place selection valve�
*
Device running continu-
ously
Float sensor blocked in water
start position
i
Clean the float� *
i
Free up the float sensor linkage
so that it can move freely� *
Start signal at the signal input (X2)
i
Check the control voltage� *
Waster water line/siphon trap
dirty
i
Clean the waste water line/si-
phon trap� *
Noise at the spittoon
Outlet ineffective or not vented
i
Check or retrofit the ventilation� *
EN
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...