Appendix
Resistance to electromagnetic interference (immunity) for all devices and systems
The device is designed for use in electromagnetic environments specified below� The customer or
operator of the device should ensure that the device is operated such an environment�
Interference immu-
nity tests
IEC 60601 - test
level
Compliance level
Electromagnetic environment
- guidelines
Electrostatic dis-
charge (ESD) in acc�
with IEC 61000-4-2
±
8 kV contact dis-
charge
±
2; 4; 8; 15 kV air
discharge
±
8 kV contact dis-
charge
±
2; 4; 8; 15 kV air
discharge
Floors should be made of wood
or cement, or covered with
ceramic tiles� If the floor is cov-
ered by synthetic material, then
the relative humidity must be at
least 30%�
Electrical fast tran-
sient/burst immunity
test in accordance
with IEC 61000-4-4
±
2 kV for mains
cables
±
1 kV for input and
output cables
±
2 kV for mains
cables
±
1 kV for input and
output cables
The quality of the supply voltage
should correspond to a typical
commercial or hospital environ-
ment�
Voltage surge in
accordance with
IEC 61000-4-5
±
0,5; 1 kV voltage
outer conduc-
tor-outer conductor
±
0�5; 1 kV symmetri-
cal voltage
The quality of the supply voltage
should correspond to a typical
commercial or hospital environ-
ment�
Voltage drops, short-
term interruptions
and fluctuations of
the supply voltage
in accordance with
IEC 61000-4-11
< 5% U
T
(> 95%
drop in U
T
) for 1/2
period
70% U
T
(30% drop
in U
T
) for 25 periods
< 5% U
T
(> 95%
drop in U
T
) for 5 s
< 5% U
T
(> 95%
drop in U
T
) for 1/2
period
70% U
T
(30% drop
in U
T
) for 25 periods
< 5% U
T
(> 95%
drop in U
T
) for 5 s
The quality of the supply voltage
should correspond to a typical
commercial or hospital envi-
ronment� If the operator of the
device needs the unit to con-
tinue working even if the mains
power supply is interrupted,
we recommend powering the
device from an uninterruptible
power supply (UPS) or from a
battery�
Magnetic field for
a supply frequency
(50/60 Hz) in
accordance with
IEC 61000-4-8
30 A/m
30 A/m
The magnetic fields at electrical
frequency should be within the
range of typical values encoun-
tered in a commercial or hospi-
tal environment�
Table 1: Resistance to electromagnetic interference (immunity) for all devices and systems
EN
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...