Electromagnetic interference immunity for devices or systems that are not life-sustaining
Portable and mobile communication devices should not be used any closer to the unit (including ca-
bles) than the recommended safety distance, which is calculated based on the applicable formula for
the transmission frequency�
Interference im-
munity tests
IEC 60601 - test
level
Compliance
level
Recommended safety distance
Conducted HF
disturbance varia-
bles in accord-
ance with
IEC 61000-4-6
3 V
eff
150 kHz to
80 MHz
[V
1
] V
d = [3�5 / V
1
]
⋅ √
P
d = 1�2
⋅ √
P
Emitted HF distur-
bance variables in
accordance with
IEC 61000-4-3
10 V/m 80 MHz
to 2�7 GHz
[E
1
] V/m
d = [3�5 / E
1
]
⋅ √
P for 80 MHz to 800
MHz
d = 1�2
⋅ √
P for 80 MHz to 800 MHz
d = [7 / E
1
]
⋅ √
P for 800 MHz to 2�7
GHz
d = 2�3
⋅ √
P for 800 MHz to 2�7 GHz
Table 2: Electromagnetic interference immunity for units or systems that are operated in health-
care facilities
P
Rated power of the transmitter in Watts (W) in accordance with the specifications of the
transmitter manufacturer
d
Recommended safety distance in metres (m)
The field strength of stationary communication devices should be lower than the compli-
ance level for all frequencies based on inspections on site
a
�
b
Interference is possible in the environment of units that have the following symbols�
Comment 1
The higher frequency range applies for 80 MHz and 800 MHz�
Comment 2
These guidelines may not apply in all cases� The propagation of electromagnetic
radiation is affected by absorption and reflection on the building, objects and
people�
a
The field strength of stationary transmitters, such as the base stations of mobile phones and land
mobile radios, amateur radio stations, AM and FM radio and television broadcasters, for example,
cannot be accurately predicted� In order to determine the electromagnetic environment with regard to
stationary transmitters, a study of electromagnetic phenomena at the site should be considered� If the
measured field strength at the location where the unit is used exceeds the compliance levels stated
above, the unit should be monitored to verify that it works as intended� If unusual performance char-
acteristics are observed additional measures may be required, such as a changing the orientation of
the unit or moving it to a different location�
b
Over the frequency range of 150 kHz to 80 MHz, the field strength should be less than [V
1
] V/m�
EN
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...