Siringa dentale SYR3
SAFETY RULES
The SYR3 dental syringe is designed exclusively for use in dentistry, by dental professionals or an
assistant under a dentist’s supervision, in order to perform specific cleaning and drying operations
on the operating area. The device must be installed on a specific supply system (dental unit) and
be connected to a dedicated FARO hose. The device must be sterilised in an autoclave before it
is used on another patient. The device is not supplied sterile and must therefore be sterilised prior
to use. Perform only the operations contained in the following manual. In all other scases request
technical assistance.
HOSE AND SYRINGE INSTALLATION
Installation must be performed by specialised personnel only.
Check that the package contains the following components:
- Dental syringe.
- Supply hose.
- O-rings.
- Probe.
- Instruction manual.
Connect the BLUE tube to the AIR intake, the GREEN tube to the H
2
O intake and fasten the
safety cord.
Connect the hose to the syringe by tightening the ring nut. Check the water and air pressure va-
lues and make sure the fluids are at ambient temperature (Tmax < 41°C).
INSTRUCTIONS FOR USE
- For water, press the green button.
- For air, press the blue button.
- For spray, press both buttons simultaneously.
Disassembling sheath and nozzle
A
B
- To remove the she-
ath, press button
“A”
and pull off in the
direction of the arrow.
- To remove the
nozzle, unscrew ring
“B”
and pull off in
the direction of the
arrow.
- Before reassembling the syringe, grease the two seals on the inner body and the metal
tube of the nozzle with petroleum jelly.
Cleaning the device
Clean the sheath and the nozzle under running water or use alcohol-based solutions.
Sterilisation
- Pull the sheath, complete with the nozzle, from the body of the syringe.
- Clean sheath and nozzle.
- Sterilise the sheath complete with the nozzle in an autoclave. The sheath has a valve that
prevents contamination of the body of the syringe.
Cleaning nozzle holes
- If the tiny holes in the nozzle that the water and air come out of are plugged, free the by using the
specially provided probe.
GB
4
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...