7117100018L30 1712V002
15
Installation
A Place selection valve
B Rinsing unit
C Suction unit relay (alternative)
D Display panel, external
7.4 Electrical connections
Station selection valve / safety valve
i
Connect the station selection valve / safety
valve using a 2-core wire with connector to
the X3 connection of the control�
Rinsing unit
i
Connect the rinsing unit using a 2-core wire
with connector to the X3 connection of the
control�
At the connection for the rinsing unit, a
suction unit relay, for example, can be
connected if there is no isolation pres-
ent between the suction unit signal and
station selection valve in the treatment
unit� Note the power consumption of the
suction unit relay�
Display panel
The display panel is used to indicate
messages acoustically and visually (via
LEDs)�
A display panel is already integrated in the unit
and should be visible/audible at all times�
If the display panel is not visible/audible, fit an
additional display panel in an easily visible loca-
tion� The display panel is connected to the X6
socket (RJ-45 socket)� An existing Dürr Dental
display panel with a 6-pin connector can be
connected to the X6A connector when replacing
an older device�
If the installation of the amalgam separator in a
neighbouring room or in the basement results
in distances of more than 3 m, we recommend
installing a shielded network cable with RJ-45
sockets�
7.3 Electrical connections,
controller
Power supply:
– Safety transformer order number:
9000-150-46
or
– Safety transformer 24 V AC with a with an
isolator consisting of two means of patient
protection (MOPP) between the mains circuit
and secondary circuit, min� 100 VA, second-
ary fuse T 4 AH (or IEC 60127-2/V T 4 AH,
250 V)
X8
X9
X1
X2
X8 X6
X6A
X7
8
10
X3.1
H2
H1
H3
H4
X3
M
F1
24V
230V 24V
A
B
C
D
X3.2
X4
X1 Power supply in accordance with
EN 60601-1, 24 V AC
X2 Signal input 24 V AC/DC
X3�1 Place selection valve / safety valve (only
CAS 1, max� output 8 W)
X3�2 Rinsing unit (CAS 1 only)
X4 CAN bus
X6 Display panel, external (X6A = connection
for predecessor model)
X7 Sensor technology
X8 Production interface
X9 Motor
H1 Motor control display
H2 Manifold control display
H3 Place selection valve control display
H4 Control display, collecting container missing
EN
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...