9000-606-39/30 1709V004
3
Important information
2.3 Improper use
Any other usage or usage beyond this scope is
deemed to be improper� The manufacturer ac-
cepts no liability for damages resulting from this�
In these cases the user/operator will bear the
sole risk�
The Combi-Sepamatic must only be used to
process liquids emerging from the oral cavity
and not any other substances such as dust,
sludge, plaster or similar�
Only chemicals and disinfectants that will not
damage the materials, e�g� OrotolPlus or equiv-
alent, must be used�
The unit is not suitable for installation down-
stream of 2 or more workplaces� The max� wa-
ter volume of 2�0l/min must not be exceeded�
The device must not be installed with the drain
higher than the connection piece on the device�
Do not use any risers� All pipes must have a
downward gradient�
Not suitable for wet rooms� Do not use this de-
vice to aspirate flammable or explosive mixtures�
Do not use the unit in a potentially explosive en-
vironment�
2.4 Systems, connection with other
devices
Additional devices connected with medical elec-
trical devices must be proven to conform with
their corresponding IEC or ISO standards� All
configurations must continue to comply with the
standard requirements for medical systems (see
IEC 60601-1-1 or section 16 of the 3rd edition
of IEC 60601-1 respectively)�
Whoever connects additional devices to medical
electrical devices automatically becomes the
system configurator and is responsible for en-
suring that the system corresponds with the
standard requirements for systems� Local laws
take priority over the requirements outlined
above�
2 Safety
Dürr Dental has designed and constructed this
device so that when used properly and for the
intended purpose there is no danger to people
or property� Nevertheless, residual risks can re-
main� You should therefore observe the follow-
ing notes�
2.1 Intended purpose
The CS 1 Combi-Sepamatic is designed for the
continuous separation of liquids in the suction
flow of dental treatment units in dry suction sys-
tems�
2.2 Intended use
Installation in accordance with the requirements
of the water authorities in the German Federal
States or in accordance with local regulations�
Installation in dental treatment units and in prac-
tice rooms (housing version)�
Positioned in the suction line of a dry suction
system after the spittoon and manifold�
The CS 1 Combi-Sepamatic is designed for the
continuous separation of liquids in the suction
flow of an individual dental treatment unit in a
dry suction system�
The The CS 1 Combi-Sepamatic can be operat-
ed in continuous operation�
The permissible volume of supply water
amounts to min� 0�1l/min, but must not exceed
2�0l/min�
A rinsing unit with fresh water can be installed
upstream of the the CS 1 Combi-Sepamatic�
For surgical treatment and when using
the Airflow, the CS 1 Combi-Sepamatic
requires a rinsing unit to be installed,
which feeds a small amount of water to
the device during aspiration� This thins
any liquid (e�g� saliva, blood) that occurs
so it can be transported more easily�
Installation, servicing and repairs must only be
performed by qualified personnel specifically ap-
proved and authorized by Dürr Dental�
The CS 1 Combi-Sepamatic is unable to sepa-
rate amalgam� The treatment of waste water
containing amalgam requires the connection of
an amalgam separator to the CS 1 Com-
bi-Sepamatic�
EN
Summary of Contents for L2-S300
Page 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Page 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Page 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 110: ......
Page 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Page 157: ......
Page 160: ...Instructions for Use Foot control S NW S N2 S N1...
Page 191: ...32 Manufacturer s declaration...
Page 192: ...33 Manufacturer s declaration...
Page 193: ...34 Manufacturer s declaration...
Page 194: ...35 Manufacturer s declaration...
Page 196: ...1 Minilight Installation and use manual...
Page 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 223: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 251: ......
Page 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 284: ......
Page 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 287: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 314: ......
Page 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...