Analog Peak Capture:
Positive analog peak with minimum full-scale input pulse
duration of 15 msec (to 1% of full-scale accuracy), 25 msec (to 0.1% of full-scale
accuracy), and 35 msec (to 0.02% of full-scale accuracy); digitally held for indefi-
nite display; PEAK and TRACK modes controlled by front-panel button, rear-panel
logic input, or computer-port command
Internal 15-Segment Linearization
, programmable from front panel
Analog Output:
Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input. May be sourced by conditioned input (A) after the FIXED ANALOG
FILTER; (B) after PEAK CAPTURE; or (C) after the SELECTABLE ANALOG FILTER
(see Fig. 15, Section 2.d)
Frequency Characteristics and Step-Response Settling Times:
AFTER THE FIXED ANALOG FILTER and AFTER +PEAK IN "PEAK" or
"TRACK" MODE:
3 dB down at 100 Hz; 60 dB down at 750 Hz. Settling Time
to 1% of final value: 15 msec; to 0.1% of final value: 25 msec; to 0.02% of final
value: 35 msec
AFTER THE SELECTABLE ANALOG FILTER:
See "Analog Filtering for Dis-
played Reading" for the Model 3530, above
Configuration:
Single-ended, return to SYSTEM COMMON
Full-Scale Range:
±5 V nominal, ±8 V maximum
Offset Range Adjustment:
±25.5% in 0.1% increments
Allowable Loading:
5 mA, max.
Offset Accuracy:
0.1% max
Span Accuracy:
0.2% max
Offset and Span Drift:
±50 ppm/°C; ±20 ppm/month
A.2.c M
ODEL
3540 F
REQUENCY
I
NPUT
C
ONDITIONER
Input Type:
Any AC or unipolar pulse signal, grounded or floating, irrespective of
waveform
Frequency Range:
From 10% to 100% of 250, 500, 1000, 2000, 4000, 8000, 16000,
or 32000 Hz
Threshold Level:
Selectable to accommodate signals from 100 mV to 100 V
Excitation (for "Zero-Velocity" sensors):
Nominal 10 (i.e., ±5) V-DC ± 5%; ±50 mA
maximum
Normal-Mode Range:
±100 V operating, ±150 V without instrument damage
Common-Mode Range:
±100 V operating and without instrument damage
Common-Mode Rejection Ratio:
-120 dB at 60 Hz; -60 dB at 1 kHz
Input Impedance:
Normal-Mode: 200 k
Ω
; Common-Mode: 50 k
Ω
Offset:
Initial: ±0.05% of full scale; vs. Temperature: ±25 ppm/°C; vs. Time: ±20
ppm/month
Gain Accuracy:
±0.02% of full scale
Gain Stability:
Vs. Temperature: ±40 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading:
5-pole filter with selectable low-pass
corner frequency of 2.5, 5, or 10 Hz
A.4
A
3500 S
ERIES
S
PECIFICATIONS
A.2 I
NDIVIDUAL
C
ONDITIONER
S
PECIFICATIONS
: M
ODEL
3540
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...