CAL
CALIBRATION
CAL=MXB [CMT]
Sets the active calibration method to "MXB" (i.e., "TWO-POINT" or "CALCULATED" cali-
bration using the "y = mx+b" linear equation). For RS-485 (only), returns
ACK [EOT]
.
CAL=LIN [CMT]
Sets the active calibration method to "15-SEGMENT LINEARIZATION." Not recognized
by the
Model 3510 Thermocouple Conditioner
. For RS-485 (only), returns
ACK
[EOT]
.
CAL [CMT]
Reads the currently active calibration method. Returns
MXB [EOT]
or
LIN [EOT]
,
except for the
Model 3510 Thermocouple Conditioner
, which can only return
MXB
[EOT]
.
CHN
CHANNEL
CHN [CMT]
Reads the current measurement data. Outputs
LBLn,w,sEUS [EOT]
, where LBL =
header string (optional—see
LABEL (LBL)
command); n = node number (optional—
see
ECHO (ECO)
command); w = data value; s = limit-status indicator number (option-
al—see
LIMITS (LIM)
command); EUS = tailer string (optional—see
ENGINEERING
UNIT STRING (EUS)
command).
CMT
COMMAND TERMINATOR
CMT=c [CMT]
Specifies the single ASCII CONTROL CHARACTER c to be recognized as the INPUT
(or "COMMAND") TERMINATOR for all commands received via the RS-232/485 Inter-
face Port; c must be entered as a hexadecimal byte in square brackets ([01] through
[1F]).
CMT
must be the same for all network nodes, and cannot be the ASCII
[ESC]
character. For RS-485 (only), returns
ACK [EOT]
.
CMT [CMT]
Reads the currently effective COMMAND TERMINATOR. Returns
c [EOT]
.
DMP DUMP
DMP [CMT]
Same as
CHANNEL (CHN)
command, above.
ECO
ECHO
ECO=ON [CMT]
Causes the 3500 instrument's NODE NUMBER to be included in all transmissions from
the RS-232/485 Interface Port in response to a
CHANNEL (CHN)
or
DUMP (DMP)
command. For RS-485 (only), returns
ACK [EOT]
.
ECO=OFF [CMT]
Cancels Node-Number "echo" for
CHN
and
DMP
transmissions. For RS-485 (only),
returns
ACK [EOT]
.
ECO [CMT]
Reads current Node-Number "echo" status. Returns
ON [EOT]
or
OFF [EOT]
.
EMM SCALING FACTOR
EMM=m [CMT]
Sets to m the SCALING FACTOR to be applied to the data reading, where -32700
≤
m
≤
32700. Also sets desired precision (decimal-point location). For RS-485 (only),
returns
ACK [EOT]
.
IMPORTANT:
THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION— THAT IS,
WHEN THE
CAL=MXB [CMT]
COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT]
, UNPREDICTABLE READINGS COULD OCCUR.
EMM [CMT]
Reads current SCALING FACTOR. Returns
m [EOT]
.
B.4
B
C
OMMAND AND
R
ESPONSE
S
YNTAX
B.8 3500 M
NEMONIC
C
OMMANDS
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...