Analog Output:
Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input.* May be sourced by conditioned input (A) after the FIXED ANA-
LOG FILTER; (B) after PEAK CAPTURE; or (C) after the SELECTABLE ANALOG FIL-
TER (see Fig. 15, Section 2.d)
Frequency Characteristics and Step-Response Settling Times:
AFTER THE FIXED ANALOG FILTER or AFTER +PEAK IN "TRACK" MODE:
3 dB down at 2 KHz; 60 dB down at 16 kHz. Settling Time to 1% of final value:
0.6 msec; to 0.1% of final value: 0.75 msec; to 0.02% of final value: 1.25 msec
AFTER +PEAK IN "PEAK" MODE:
Settling Time to 1% of final value: 6 msec; to
0.1% of final value: 12 msec; to 0.02% of final value: 20 msec
AFTER THE SELECTABLE ANALOG FILTER:
See "Analog Filtering for Dis-
played Reading" for the Model 3570, above
Configuration, Full-Scale Range, Offset Range Adjustment, Allowable
Loading, Offset Accuracy, Span Accuracy, Offset and Span Drift:
See the
Model 3530 LVDT Conditioner, above
A.2.f M
ODEL
3578 AC S
TRAIN
G
AGE
C
ONDITIONER
Transducer Types and Ranges:
Conventional 4-arm strain gage bridge, nominal
90
Ω
or higher, with a full-scale range of 0.75, 1.50, or 3.00 mV/V. 1/4-, 1/2-, or full-
bridge gage configurations can be accommodated by appropriate external
bridge-completion circuitry supplied by the user. Since channel zeroing is by dig-
ital techniques, no input balance control is provided. The allowable input range,
therefore, must include any initial unbalance (which, in commercially produced
strain gage transducers, is negligible). Other transducers may have to be exter-
nally trimmed to be used with the instrument, if zero unbalance exceeds 20% of
full scale.
Actual
Absolute
Input at which
Normal-Mode
Maximum
Display Reading
Nominal
Input-Signal
Overrange
Corresponds
Input Range
Voltage (Max)
(33%)
to "m" Term
3.00 mV/V
13.5 mV
4.0 mV/V
4.0 mV/V
1.50 mV/V
6.75 mV
2.0 mV/V
2.0 mV/V
0.75 mV/V
3.375 mV
1.0 mV/V
1.0 mV/V
AC Excitation:
Regulated 3 V-AC (rms) at 3280 Hz; 50 mA (rms), maximum
Amplifier:
AC-coupled demodulator with user-settable phase and symmetry con-
trols
Normal-Mode Range:
±12 mV rms, operating; ±8 V-DC without instrument dam-
age
Common-Mode Range:
±0.8 V-DC operating, ±8 V-DC without instrument damage
Common-Mode Rejection Ratio:
infinite at DC; -90 dB at 60 Hz; -80 dB at 1 kHz;
-60 dB at 3 kHz
Input Impedance (Differential and Common-Mode):
10 M
Ω
Offset:
Initial: ±3% of full scale; vs. Temperature: ±0.005% f.s./°C; vs. Time: ±0.002%
f.s./month
A.7
3500 S
ERIES
S
PECIFICATIONS
A
A.2 I
NDIVIDUAL
C
ONDITIONER
S
PECIFICATIONS
: M
ODEL
3578
* When (1) the analog output scale is 100.0%, (2) the analog output offset is 0.0%, and (3) the
excitation is 10V, then any of the following inputs will result in an analog output of 5 V: an input
of 4 mV/V on the 3.0-mV/V range; an input of 2 mV/V on the 1.5-mV/V range; or an input of 1
mV/V on the 0.75-mV/V range.
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...