3.
When the desired voltage range appears, press
ENTER
to return to the
"RANGE" display. Then press
SETUP
to return to the "Setup" display, and pro-
ceed to Section 3.a.6.
3.a.5.e S
ETTING THE
M
ODEL
3570’
S
DC S
TRAIN
G
AGE
I
NPUT
R
ANGE AND
E
XCITATION
The
Model 3570 DC Strain Gage Conditioner
must be set to match the full-
scale range of the strain gage transducer.
1.
If, after pressing the
RANGE
button, you want to display the current input-
range setting, press
ENTER
. Otherwise, press
RANGE
again and proceed to
Step 5. If you press
ENTER
, the unit will display a number representing the
current input-range setting (in mV/V): ".75," "1.5," or "3.0."
2.
To change the input range, you can cycle forwards or backwards through the
allowed numbers by pressing the top or bottom segment, respectively, of any
digit display (even if it is blank).
4.
When the desired mV/V range appears, press
ENTER
. The unit will display
"ECITE" (for "EXCITATION"):
5.
If you want to display the current excitation setting, press
ENTER
. Otherwise,
press
RANGE
again and proceed to Step 7. If, for example, the conditioner is
set for an excitation of 5 V-DC, when you press
ENTER
it will display "5."
6.
To change the excitation level, you can cycle forwards or backwards through
the allowed settings (2, 5, and 10 V) by pressing the top or bottom segment,
respectively, of any digit display (even if it is blank).
7.
When the desired excitation appears, press
ENTER
to return to the "RANGE"
display. Then press
SETUP
to return to the "Setup" display, and proceed to
Section 3.a.6.
3.a.5.f S
ETTING THE
M
ODEL
3578’
S
AC S
TRAIN
G
AGE
I
NPUT
R
ANGE
The
Model 3578 AC Strain Gage Conditioner
must be set to match the full-
scale range of the strain gage transducer.
1.
After pressing the
RANGE
button, press
ENTER
. The current range setting
will now appear. If, for example, the conditioner is set for a full-scale input of
1.5 mV/V, the display will show the number "1.5."
2.
To change the input range, you can cycle forwards or backwards through the
allowed settings (.75, 1.5, and 3.0 mV/V) by pressing the top or bottom seg-
ment, respectively, of any digit display (even if it is blank).
3.
When the desired mV/V range appears, press
ENTER
to return to the
"RANGE" display. Then press
SETUP
to return to the "Setup" display.
3.11
S
ETUP
: I
NSTRUMENT
C
ONFIGURATION
3
3.a C
ONFIGURING
T
HROUGH THE
F
RONT
P
ANEL
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...