P
HASE
/S
YMMETRY
A
DJUSTMENT FOR A
L
EBOW
1800 S
ERIES
T
RANSDUCER
1.
WHEN USING THE MODEL 3578 WITH A
LEBOW 1800 SERIES TRANSDUCER
,
YOU SHOULD FIRST INSTALL THE SHUNT RESISTOR SUPPLIED WITH THE
TRANSDUCER. Refer to Fig. 11(c) and connect the resistor between Pin 5 of the
CONDITIONER CONNECTOR and the transducer's +SIGNAL line (you will not use
the "CAL SENSE" pin (Pin 4) in this case).
2.
Locate the "CAL/RUN" Switch in the cable harness of the 1800 Series transducer.
Place this switch in the "CAL" position.
3.
Establish a zero input by removing all load from the 1800 Series transducer.
4.
Enter SETUP MODE (see Section 3.a.1), and press the
CAL
button. If the 3578 unit
does not now display "
2 Pt
," continue to press
CAL
until it does.
5.
Press the
ENTER
button. The unit will display "OFSET" (for "ZERO OFFSET").
6.
Press the
CAL
button. The unit will display "FORCE."
7.
Press
ENTER
. The unit will display "SHP" (for "SHUNT CALIBRATION—POSITIVE").
8.
Press
ENTER
again. The display will show the status of the positive shunt: "ON"
means that the shunt is connected; "OFF," that it is not. If the positive shunt is cur-
rently
OFF
, turn it
ON
by pressing the top or bottom segment of any displayed
character, followed by
ENTER
. The unit will display "SHN" (for "SHUNT CALIBRA-
TION—NEGATIVE").
9.
Press
ENTER
. The display will show the status of the negative shunt (it should be
OFF
, since you just turned
ON
the positive shunt). Press
ENTER
. This will return
you temporarily to a display of the "live" analog input.
10. Now adjust the rear-panel PHASE ADJUSTMENT CONTROL (see Fig. 20) until a
maximum reading is obtained.
11. Press the
CAL
button. The display will return to "OFSET" (the positive shunt has
been automatically turned
OFF
).
12. Press
ENTER
. This will return you temporarily to a display of the "live" analog
input.
13. Press ANY KEY EXCEPT
CAL
OR
SETUP
. This will freeze the displayed data
value.
14. Using the NUMERIC BUTTONS as explained in Section 1.e, change the displayed
value to "0" (zero).
15. Press
ENTER
. The display will return to "OFSET." Now press
CAL
. The unit will
again display "FORCE."
16. Repeat Steps 7 through 9.
17. Press ANY KEY EXCEPT
CAL
OR
SETUP
. This will freeze the displayed data
value.
4.6
4
S
ETUP
: I
NSTRUMENT
C
ALIBRATION
4.b P
HASE AND
S
YMMETRY
A
DJUSTMENT OF THE
M
ODEL
3578
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...