5.a
RS-232/485 C
OMMUNICATIONS
5.a.1 RS-232/485 O
UTPUTS
:
CHN
AND
DMP C
OMMANDS
A 3500 Series instrument will issue output from its RS-232/485 Interface Port in
response to
• a command to "PRINT" from the front-panel
button or the rear logic I/O
connector (when the interface is in RS-232 mode);
• a
CHANNEL (CHN)
or
DUMP (DMP)
command received through the Inter-
face Port (in either RS-232 or RS-485 mode);
• a valid interrogation command received through that port (in RS-232 mode),
or
• any valid or invalid command received through that port (in RS-485 mode).
For the initiation of RS-232 PRINT transmissions, see Section 5.d, below. For pos-
sible responses to mnemonic commands, see Appendix B.
The general form of the
CHANNEL (CHN)
command is
CHN [CMT]
where
[CMT]
is the "COMMAND TERMINATOR" (that is, the INPUT TERMINATOR
character which the 3500 instrument has been set to recognize—see Section
3.a.3). This command instructs the unit to transmit the current data for its single
measurement channel from the RS-232/485 Interface Port, in the format selected
by the user.
The general form of the
DUMP (DMP)
command is
DMP [CMT]
This command is functionally identical to the
CHN
command.
5.a.2 D
ATA
-T
RANSMISSION
F
ORMAT
The data transmitted when the 3500 instrument is instructed to "PRINT" (in RS-232
mode) or when a
CHANNEL (CHN)
or
DUMP (DMP)
command has been
received through the RS-232/485 Interface Port (in either RS-232 or RS-485
mode) can include
a.
An optional "HEADER" string defined by a
LABEL (LBL)
command of
LBL=$ [CMT]
where "$" is a string of up to eight characters, including spaces.* The data
"label" is typically a channel name or description (e.g., CHAN 4, INPUT 7, TEST
R, etc.). It has no numeric effect on measurement scaling.
b.
An optional NODE-NUMBER "ECHO," if so indicated by the user during the
button setup procedure (Section 3.a.10) or by application of an
ECHO
(ECO)
command of
ECO=ON [CMT]
5.1
R
UN
-T
IME
O
PERATION
5
5.a RS-232/485 C
OMMUNICATIONS
* To remove an existing
LBL
string from the transmission, just command
LBL=N/A [CMT].
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...