VER
VERSION
VER [CMT]
Causes "version" information (Model No. and Software Version No.) to be transmitted
from the RS-232/485 Interface Port, followed by
[EOT]
.
ZRO
ZERO
ZRO=v [CMT]
Sets the ZERO OFFSET ("b" term) so that the existing input yields a reading equal to v
(-32700
≤
v
≤
32700). In "TWO-POINT" calibration, must precede the
FORCE (FRC)
command. For RS-485 (only), returns
ACK [EOT]
.
IMPORTANT:
THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION— THAT IS,
WHEN THE
CAL=MXB [CMT]
COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT]
, UNPREDICTABLE READINGS COULD OCCUR.
[
NOTE:
There is no valid interrogation form for the
ZRO
command.]
B.11
C
OMMAND AND
R
ESPONSE
S
YNTAX
B
B.8 3500 M
NEMONIC
C
OMMANDS