B.7
C
OMMANDS
T
HAT
I
NITIATE
D
ATA
T
RANSMISSIONS
The
CHANNEL (CHN)
and
DUMP (DMP)
commands, discussed in Section 5.a.1, are
the only ones that belong to this category—with the possible addition of
VERSION
(VER)
, which causes a transmission of instrument "version" information to be issued
from the RS-23/485 Interface Port.
B.8
T
ABLE OF
3500 M
NEMONIC
C
OMMANDS
NOTE: Unless otherwise indicated, every command in the following list
applies equally to all 3500 Series conditioner models.
For the use of CONFIGURATION SETUP commands, see also Section 3.b. For
the use of CALIBRATION SETUP commands, see also Section 4.d. For the use
of RUN-TIME commands, see also Section 5.
AOG
ANALOG OUTPUT GAIN
AOG=v [CMT]
Sets to v the ANALOG OUTPUT GAIN, where v is a percentage of full scale to the near-
est tenth of a percent between 74.4% and 125.6%. For RS-485 (only), returns
ACK
[EOT]
.
AOG [CMT]
Reads current ANALOG OUTPUT GAIN. Returns
v [EOT]
.
AOO
ANALOG OUTPUT OFFSET
AOO=v [CMT]
Sets to v the ANALOG OUTPUT OFFSET, where v is a percentage of full scale to the
nearest tenth of a percent between -25.6% and 25.6%. For RS-485 (only), returns
ACK
[EOT]
.
AOO [CMT]
Reads current ANALOG OUTPUT OFFSET. Returns
v [EOT]
.
BAU
BAUD RATE
BAU=b,s,d,p [CMT]
Sets BAUD RATE and other protocols for the RS-232/485 Interface Port: b = Baud rate:
1 = 300, 2 = 600, 3 = 1200, 4 = 2400, 5 = 4800, 6 = 9600, 7 = 19.2K, 8 = 38.4K, 9 = 153.6K;
s = No. of Stop Bits: 1 or 2; d = No. of Data Bits: 7 or 8; p = Parity: E = EVEN, O = ODD,
N = NONE, Z = ZERO. For RS-485 (only), returns
ACK [EOT]
.
BAU [CMT]
Reads current protocol values for RS-232/485 Interface Port. Returns
b,s,d,p [EOT]
.
BEE
ZERO OFFSET
BEE=b [CMT]
Sets to b the ZERO OFFSET to be applied to the data reading, where -32700
≤
b
≤
32700. For RS-485 (only), returns
ACK [EOT]
.
IMPORTANT:
THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—THAT IS,
WHEN THE
CAL=MXB [CMT]
COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT]
, UNPREDICTABLE READINGS COULD OCCUR.
BEE [CMT]
Reads current ZERO OFFSET. Returns
b [EOT]
.
B.3
C
OMMAND AND
R
ESPONSE
S
YNTAX
B
B.7 C
OMMANDS
T
HAT
I
NITIATE
D
ATA
T
RANSMISSIONS
B.8 3500 M
NEMONIC
C
OMMANDS
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...