EQUIVALENT INPUT can be approximated from a knowledge of the Shunt Cali-
bration Resistance (R); the transducer's Bridge Resistance (B); and the transduc-
er's Full-Scale Sensitivity (K, in mV/V, full scale).
To determine the EQUIVALENT INPUT (X) as an approximate percentage of full-
scale output, you may use the following equation:
a.
Perform Steps a through f of the "Actual" two-point calibration procedure
given above.
b.
Press
ENTER
. The unit will display "SHP" (for SHUNT CALIBRATION—POSI-
TIVE).
c.
If you want to switch in the shunt resistor for a positive up-scale reading, press
ENTER
once again and proceed to Step d. If you want to switch in the shunt
resistor for a negative up-scale reading, press the
CAL
button to display
"SHN" (for "SHUNT CALIBRATION—NEGATIVE"), and press
ENTER
.
d.
After you press
ENTER
, the 3570/3578 will show the status of the selected
shunt: "ON" means that the shunt is connected; "OFF," that it is not. If the
shunt is currently
OFF
, turn it
ON
by pressing the top or bottom segment of
any displayed character, followed by
ENTER
. This will return you temporarily
to a display of the "live" analog input.
NOTE: By turning
ON
either of the two shunts, you will automatically turn
OFF
the
other one (if it is presently ON).
ALSO NOTE: Toggling the "SHP" or "SHN" display to
ON
is equivalent to entering a
command of
SHP=ON [CMT]
or
SHN=ON [CMT]
, respectively, through the RS-
232/485 Interface. Toggling a shunt to
OFF
is equivalent to entering a command
of
SHP=OFF [CMT]
or
SHN=OFF [CMT]
. As explained in Section 1.f, the state of
the instrument's positive and negative calibration shunts can also be controlled
by means of logic inputs to the rear-panel Analog Input Connector. However, if
the unit receives contradictory shunt commands from the keypad/interface and
from a shunt-control logic input, it will return to normal "live" measurement.
e.
Now freeze the displayed data value by pressing ANY KEY EXCEPT
CAL
OR
SETUP
.
f.
Using the NUMERIC BUTTONS, change the displayed value to the known
EQUIVALENT INPUT produced by the shunt, with appropriate polarity. As
before, this second calibration entry also sets the desired measurement pre-
cision.
g.
Press
ENTER
to establish your second calibration point. The display will
return to "FORCE."
h.
Press the
CAL
button. The display will return to "OFSET." Note that pressing
the
CAL
button will automatically turn
OFF
whichever shunt is presently ON,
so that normal measurement will resume as soon as you return the condition-
er to "RUN" mode.
i.
To exit the
CAL
button sequence and return to the "Setup" display, press
SETUP
.
4.11
S
ETUP
: I
NSTRUMENT
C
ALIBRATION
4
4.c C
ALIBRATING
T
HROUGH THE
F
RONT
P
ANEL
X = %
25000(mV/V)B
KR
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...