Conditioner
"r"
Actual Input
Model
value
Range or Scale
3560 Voltage Conditioner
0.5V
±0.5 V-DC
1V
±1.0 V-DC
2V
±2.0 V-DC
5V
±5.0 V-DC
10V
±10.0 V-DC
20V
±20.0 V-DC
3570 DC Strain Gage Conditioner
.75
0.75 mV/V
1.5
1.5 mV/V
3.0
3.0 mV/V
3578 AC Strain Gage Conditioner
.75
0.75 mV/V
1.5
1.50 mV/V
3.0
3.00 mV/V
3.b.5 E
XCITATION
(M
ODELS
3570
AND
3578 ONLY)
EXCITATION (EXC): EXC=v [CMT]
Sets the excitation level to "v," as given by the following table:
Excitation No. ("v")
Excitation
2
2 V (±1 V)
5
5 V (±2.5 V)
10
10 V (±5 V)
3.b.6 S
ENSITIVITY
(M
ODEL
3540 ONLY)
SENSITIVITY (SEN): SEN=s [CMT]
Sets the input sensitivity range to "s," as given by the following table:
Sensitivity No. ("s")
Sensitivity Range
.1
0.1 to 5 V
1
1 to 20 V
10
10 to 100 V
3.b.7 F
ILTER
RESPONSE (RES): RES=f [CMT]
Sets the analog filter corner to "f" Hz. For the
Model 3540 Frequency Condi-
tioner
, "f" = 2 (for 2.5), 5, or 10; for all other conditioners except the Thermo-
couple Conditioner, "f" = 5, 10, or 20. For the
Model 3510 Thermocouple
Conditioner
, the
RES
command is inactive— although an interrogation of
RES [CMT]
will always answer "20."
FILTER (FIL): FIL=f [CMT]
Sets the digital filtering constant to "f" (where 0
≤
f
≤
9).
3.22
3
S
ETUP
: I
NSTRUMENT
C
ONFIGURATION
3.b C
ONFIGURING
T
HROUGH THE
RS-232/485 I
NTERFACE
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...