COMMAND TERMINATOR (CMT): CMT=c [CMT]
Sets the COMMAND (or "INPUT") TERMINATOR to "c" (where c is a single
ASCII character entered as a hexadecimal byte in square brackets, [01]
through [1F]).
END-OF-TRANSMISSION TERMINATOR (EOT): EOT=$ [CMT]
Sets the END-OF-TRANSMISSION (or "OUTPUT") TERMINATOR to "$" (where
$ is a string of up to four ASCII characters, each entered as a hexadecimal
byte in square brackets, [01] through [1F]). In a multinode network, the last
EOT character must be the same as the CMT character.
NOTE:
A working EOT cannot include a character of "[00]," except as a default
value for unused characters. If, for example, you specify a two-character EOT
of "[0A][0D]" and then use the front-panel procedure given in Section 3.a to
interrogate for the current EOT characters, you will be told that "Eot1" = [0A];
"Eot2" = [0D]; "Eot3" = [00]; and "Eot4" = [00]. In fact, the instrument will stop
transmitting as soon as it perceives an EOT character of [00]. Thus, if you
were to enter an EOT of "[00][0D] ...," no EOT at all would be transmitted.
3.b.3 T
HERMOCOUPLE
T
YPE
(M
ODEL
3510 ONLY)
TYPE (TYP): TYP=y [CMT]
or
TYP=y,N [CMT]
The first version of the command sets the Thermocouple Type to "y" (where
"y" = E, J, K, R, S, T, or B) and enables the internal cold-junction reference.
The second version sets the Thermocouple Type to "y" and disables the inter-
nal cold-junction.
3.b.4 I
NPUT
R
ANGE OR
S
CALE
RANGE (RNG): RNG=r [CMT]
Sets the input range or scale to "r," as given by the following table:
Conditioner
"r"
Actual Input
Model
value
Range or Scale
3510 TC Conditioner
CEN
degrees C
FAR
degrees F
3530 LVDT Conditioner
5
0-150 mV/V "normal";
0-1 V/V "long-stroke"
10
0-300 mV/V "normal";
0-2 V/V "long-stroke"
20
0-600 mV/V "normal";
0-4 V/V "long-stroke
3540 Frequency Conditioner
250
250 Hz
500
500 Hz
1000
1000 Hz
2000
2000 Hz
4000
4000 Hz
8000
8000 Hz
16000
16000 Hz
32000
32000 Hz
(cont’d)
3.21
S
ETUP
: I
NSTRUMENT
C
ONFIGURATION
3
3.b C
ONFIGURING
T
HROUGH THE
RS-232/485 I
NTERFACE
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...