Most of the following ERROR CODES refer to the INTERNAL LINEARIZATION pro-
cedure discussed in Section 4.c.4.
Error Display
Meaning
Err1
Insufficient input error—i.e., the desired display resolution
is too great for the input change.
Err2
Table entry error; the entered "input" is within the end-
points of another segment.
Err3
While calculating the "b" term of the segment, an arith-
metic overrange has occurred in justifying the decimal
points.
Err4
The current range is not legal for this instrument. Re-
enter the range.
Err5
The engineering-unit "output" entered for the current seg-
ment is equal to that of the previous segment (resulting in
a slope of "0").
Err6
The "input" of the current segment is equal to that of the
previous segment.
Err7
The entered "input" has an incorrect precision.
C.1