3.b
C
ONFIGURING
T
HROUGH THE
RS-232/485 I
NTERFACE
When proper RS-232/485 communications have been established with an exter-
nal computer or terminal, it is possible for a 3500 Series instrument to receive
and respond to SETUP COMMANDS issued directly to it through this interface.
There is a specific SETUP COMMAND for each of the front-panel button proce-
dures described in the previous section, with the exception of the node-number
assignment (see the
NODE (NOD)
command, below). These commands are list-
ed below. For a full discussion of "Command and Response Syntax," along with a
list of all valid mnemonic commands, see Appendix B of this manual.
Except for
NODE (NOD)
, each of the following SETUP COMMANDS can take a
"WRITE" or "READ" form. Only the "WRITE" form is given here. Again, for full
details, see Appendix B. In the following command expressions,
[CMT]
means
"COMMAND TERMINATOR" (that is, the INPUT TERMINATOR character which the
unit has been set to recognize—see the procedure for setting the INPUT TERMI-
NATOR via the
COM
button in Section 3.a.3, above).
IMPORTANT:
WHEN CONFIGURING YOUR UNIT "REMOTELY" THROUGH THE
RS-232/485 INTERFACE, DON'T FORGET TO ISSUE A COMMAND OF
SAV=ON [CMT]
WHEN YOU'RE FINISHED. THIS COMMAND WILL SAVE THE EXISTING SETUP
CONFIGURATION TO NONVOLATILE EEPROM.* IF NOT SAVED, THE SETUP VAL-
UES YOU HAVE JUST ENTERED WILL BE LOST ON INTERRUPTION OF POWER.
3.b.1 S
ECURITY
C
ODE
SECURITY CODE (SCD): SCD=c [CMT]
Sets the security code to "c" (where 0
≤
c
≤
98).
3.b.2 C
OMMUNICATIONS
P
ARAMETERS
BAUD RATE (BAU): BAU=b,s,d,p [CMT]
Sets the baud rate ("b"), number of stop bits ("s"), number of data bits ("d"),
and parity ("p"). Refer to Appendix B for the values these variables can take.
NODE (NOD): NOD [CMT]
Reads the node number of the currently "open" node (for "opening" a node,
see Section 5.a.3). THERE IS NO "WRITE" FORM OF THIS COMMAND; A
NODE NUMBER CAN BE ASSIGNED TO A UNIT ONLY THROUGH THE
FRONT-PANEL
COM
BUTTON SEQUENCE DESCRIBED IN SECTION 3.a.3.
3.20
3
S
ETUP
: I
NSTRUMENT
C
ONFIGURATION
3.b C
ONFIGURING
T
HROUGH THE
RS-232/485 I
NTERFACE
* Note that an interrogation of the form
SAV [CMT]
will return "ON" or "OFF," depending on
whether or not the setup is in the process of being automatically saved as a result of pressing
the front-panel
SETUP
button (Section 3.a.1). Note too that the SETUP COMMAND
SAV=OFF
[CMT]
has no effect.
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...