measured phenomenon, expressed in the desired engineering units and precision).
For RS-485 (only), returns
ACK [EOT]
.
IMPORTANT:
THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3540 IS SET
FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—THAT IS, WHEN THE
CAL=MXB [CMT]
COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT]
,
UNPREDICTABLE READINGS COULD OCCUR.
FRQ [CMT]
Reads current "FRQ" calibration constants. Returns
i,u [EOT]
. See Section 4.d.4
regarding the answer returned in response to an
FRQ [CMT]
interrogation.
HHY
HIGH HYSTERESIS
HHY=p [CMT]
Sets the HIGH HYSTERESIS window depth to p (where p is a percentage of the exist-
ing SCALING FACTOR ("m") value to the nearest tenth of a percent). For RS-485 (only),
returns
ACK [EOT]
.
HHY [CMT]
Reads current HIGH HYSTERESIS window depth. Returns
p [EOT]
.
HIL
HIGH LIMIT
HIL=h [CMT]
Sets the HIGH LIMIT value to h (-32700
≤
h
≤
32700; h must be greater than or equal to
the existing
LOW LIMIT (LOL)
value). For RS-485 (only), returns
ACK [EOT]
.
HIL [CMT]
Reads current HIGH LIMIT value. Returns
h [EOT]
.
HLA
HIGH LATCH
HLA=ON [CMT]
Sets the HIGH LIMIT to LATCHING status. For RS-485 (only), returns
ACK [EOT]
.
HLA=OFF [CMT]
Sets the HIGH LIMIT to NONLATCHING status. For RS-485 (only), returns
ACK [EOT]
.
HLA [CMT]
Reads current HIGH-LIMIT LATCH status. Returns
ON [EOT]
or
OFF [EOT]
.
LBL
LABEL
LBL=$ [CMT]
Specifies "header" string $ for all transmissions from the RS-232/485 Interface Port in
response to a
CHANNEL (CHN)
or
DUMP (DMP)
command ($ is a string of up to 8
characters, including spaces). For RS-485 (only), returns
ACK [EOT]
.
LBL=N/A [CMT]
Clears all characters from the existing "header" string. For RS-485 (only), returns
ACK
[EOT]
.
LBL [CMT]
Reads current "header" string. Returns
$ [EOT]
or
N/A [EOT]
.
LFC
LINEARIZATION FORCE
LFC=V [CMT]
Sets to v the OUTPUT value for the linearization segment corresponding to the exist-
ing "live" INPUT. Not recognized by the
Model 3510 Thermocouple Conditioner
.
For RS-485 (only), returns
ACK [EOT]
.
[
NOTE:
There is no valid interrogation form for the
LFC
command.]
LHY
LOW HYSTERESIS
LHY=p [CMT]
Sets the LOW HYSTERESIS window depth to p (where p is a percentage of the exist-
ing SCALING FACTOR ("m") value to the nearest tenth of a percent). For RS-485 (only),
returns
ACK [EOT]
.
B.6
B
C
OMMAND AND
R
ESPONSE
S
YNTAX
B.8 3500 M
NEMONIC
C
OMMANDS
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...