The following front-panel buttons are active during normal instrument operation.
For full details on all standard run-time procedures—including peak capture, tare,
and printout—see Section 5.
When this button is pressed while the 3500 instrument is in RUN-TIME MODE,
the operator will be asked to enter the current SECURITY CODE (or to enter a
new code, if it is currently "0"). Once the correct security code has been entered,
the instrument will be placed in SETUP MODE, and the display shown in Fig. 1(b)
will appear. The button will remain lit as long as the instrument is in the SETUP
MODE. While in SETUP MODE, the
SETUP
button is used to exit the current
setup procedure or to exit SETUP MODE altogether.
When the "PEAK" logic input terminal is disconnected (i.e., at the "Logic 0" level),
pressing the
PEAK/TRACK
button once will place the 3500 instrument in the
"+PEAK" MODE. It will then continuously display the most positive value experi-
enced since the button was pressed (see Figs. 22 and 23, Section 5.b).* The but-
ton will remain lit as long as the instrument is in the "+PEAK" MODE. Each
captured maximum value will remain in digital memory until occurrence of a sub-
sequent higher-valued maximum or until the instrument is returned to "TRACK"
MODE. Pressing the button a second time will cause the instrument to return to
"TRACK" MODE.
When the "PEAK" logic input terminal is connected to GROUND (i.e., at the "Logic
1" level), the
PEAK/TRACK
button is DISABLED.**
NOTE: The
PEAK/TRACK
button is not active during SETUP MODE.
E.1
R
UN
-T
IME
B
UTTON
F
UNCTIONS
E
R
UN
-T
IME
B
UTTON
F
UNCTIONS
* For all models except the
Model 3510 Thermocouple Conditioner
and the
Model 3540 Fre-
quency Conditioner
, the captured "+PEAK" can be issued as the instrument's ANALOG OUT-
PUT, when the appropriate jumper connections have been made (see Section 2.d). Note also
that when the "PEAK" logic input terminal is at "Logic 0," a command of
PEK=ON [CMT]
applied via the RS-232/485 Interface Port has the same effect as pressing the
PEAK/TRACK
button the first time, while
PEK=OFF [CMT]
has the same effect as pressing the button the
second time (see Section 5.b for the use of these commands).
** Changing the logic level of the "PEAK" logic input terminal from "Logic 0" to "Logic 1" has the
same effect as pressing the
PEAK/TRACK
button the first time (i.e., it places the 3500 instru-
ment in "+PEAK" MODE), while changing the "PEAK" logic input terminal from "Logic 1" to
"Logic 0" has the same effect as pressing the button the second time (i.e., it returns the instru-
ment to "TRACK" MODE). A Logic 1 level at Logic Input 6 also disables the
PEK=ON [CMT]
and
PEK=OFF [CMT]
mnemonic commands as well as the
PEAK/TRACK
button (see Section
5.b).
SET UP
PEAK
TRACK
Appendix E
Run-Time Button Functions
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...