EOT
END-OF-TRANSMISSION TERMINATOR
EOT=$ [CMT]
Specifies an OUTPUT (or "END-OF-TRANSMISSION") TERMINATOR of $, where $ is a
string of up to four ASCII characters, each entered as a hexadecimal byte in square
brackets ([01] through [1F]—the current transmission will be instantly halted if the
instrument perceives an
EOT
character of
[00]
).
EOT
must be the same for all net-
work nodes. In a multinode network, the last
EOT
character must be the same as the
COMMAND TERMINATOR (see the
CMT
command). For RS-485 (only), returns
ACK
[EOT]
.
EOT [CMT]
Reads the currently effective END-OF-TRANSMISSION TERMINATOR. Returns
$
[EOT]
.
EUS
ENGINEERING UNIT STRING
EUS=$ [CMT]
Specifies "tailer" string $ for all transmissions from the RS-232/485 Interface Port in
response to a
CHANNEL (CHN)
or
DUMP (DMP)
command ($ is a string of up to 8
characters, including spaces). For RS-485 (only), returns
ACK [EOT]
.
EUS=N/A [CMT]
Clears all characters from the existing "tailer" string. For RS-485 (only), returns
ACK
[EOT]
.
EUS [CMT]
Reads current "tailer" string. Returns
$ [EOT]
or
N/A [EOT]
.
EXC
EXCITATION —
Model 3570 DC Strain Gage Conditioner only
EXC=v [CMT]
Sets to v volts (DC) the excitation level of the Model 3570 DC Strain Gage Conditioner
(v = 2, 5, or 10). For RS-485 (only), returns
ACK [EOT]
.
EXC [CMT]
Reads current excitation level. Returns
v [EOT]
.
FIL
FILTER
FIL=f [CMT]
Sets to f the digital filtering constant (0
≤
f
≤
9). For RS-485 (only), returns
ACK [EOT]
.
FIL [CMT]
Reads current digital filtering constant. Returns
f [EOT]
.
FRC
FORCE
FRC=v [CMT]
Sets the SCALING FACTOR ("m" coefficient) so that the existing input yields a reading
equal to v (-32700
≤
v
≤
32700). Also sets desired precision (decimal-point location).
In "TWO-POINT" calibration, must be preceded by a
ZERO (ZRO)
command. For RS-
485 (only), returns
ACK [EOT]
.
IMPORTANT:
THIS COMMAND SHOULD ONLY BE APPLIED WHEN THE 3500
INSTRUMENT IS SET FOR "TWO-POINT" OR "CALCULATED" CALIBRATION—THAT IS,
WHEN THE
CAL=MXB [CMT]
COMMAND IS IN EFFECT. IF ENTERED WHEN
CAL=LIN [CMT]
, UNPREDICTABLE READINGS COULD OCCUR.
[
NOTE:
There is no valid interrogation form for the
FRC
command.]
FRQ
FREQUENCY CALIBRATION —
Model 3540 Frequency Conditioner only
FRQ=i,u [CMT]
Sets an appropriate SCALING FACTOR for the Model 3540 Frequency Conditioner
based on "i" (the nominal full-scale rating of the frequency source—or the highest fre-
quency expected to be measured—in Hertz) and "u" (the corresponding value of the
B.5
C
OMMAND AND
R
ESPONSE
S
YNTAX
B
B.8 3500 M
NEMONIC
C
OMMANDS
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...