3.b.8 A
NALOG
O
UTPUT
ANALOG OUTPUT OFFSET (AOO): AOO=v [CMT]
Sets the analog output offset to "v" (where v is a percentage of full scale to the
nearest tenth of a percent between -25.5% and 25.5%).
ANALOG OUTPUT GAIN (AOG): AOG=v [CMT]
Sets the analog output gain to "v" (where v is a percentage of full scale to the
nearest tenth of a percent between 74.5% and 125.5%).
See Section 3.a.7 for an explanation of the special use of the
AOO
and
AOG
para-
meters for the
Model 3510 Thermocouple Conditioner
.
3.b.9 L
IMITS
HIGH LIMIT (HIL): HIL=h [CMT]
Sets the HIGH LIMIT value to "h."
HIGH HYSTERESIS (HHY): HHY=p [CMT]
Sets the HIGH HYSTERESIS window depth to "p" (where p is a percentage of
the existing SCALING FACTOR ("m") value to the nearest tenth of a percent).
HIGH LATCH (HLA): HLA=ON [CMT]
or
HLA=OFF [CMT]
Sets the HIGH-LIMIT LATCH status to "ON" or "OFF."
LOW LIMIT (LOL): LOL=L [CMT]
Sets the LOW LIMIT value to "L."
LOW HYSTERESIS (LHY): LHY=p [CMT]
Sets the LOW HYSTERESIS window depth to "p" (where p is a percentage of
the existing SCALING FACTOR ("m") value to the nearest tenth of a percent).
LOW LATCH (LLA): LLA=ON [CMT]
or
LLA=OFF [CMT]
Sets the LOW-LIMIT LATCH status to "ON" or "OFF."
3.b.10 T
ARE
O
FFSET
TARE REGISTER (TRR): TRR=v [CMT]
Sets the TARE REGISTER value to "v."
3.b.11 P
RINT AND
O
UTPUT
P
ARAMETERS
PRINT INTERVAL (PRI): PRI=t [CMT]
Sets the automatic print interval to approximately (t x 0.125) second(s), where
0
≤
t
≤
32700.
3.23
S
ETUP
: I
NSTRUMENT
C
ONFIGURATION
3
3.b C
ONFIGURING
T
HROUGH THE
RS-232/485 I
NTERFACE
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...