Common-Mode Rejection Ratio:
-116 dB at DC; -120 dB at 60 Hz
Input Impedance:
Differential: 10 megohms; Common-Mode: 0.5 megohm
Offset:
Initial: ±0.005 mV; vs. Temperature: ±0.0001 mV/°C; vs. Time: ±0.001
mV/month
Gain Accuracy:
±0.05% of absolute mV input range of -10 to +80 mV
Gain Stability:
Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading:
2-pole modified Butterworth; 3 dB down
at 1 Hz; 60 dB down at 50 Hz
Step Response Settling Time (Full-Scale Output)
To 1% of final value: 1.0 second
To 0.1% of final value: 2 seconds
To 0.02% of final value: 6 seconds
Total System Accuracy (
typical): See table, above
Digital Peak Capture:
Displays the most positive measurement reading, which is
digitally held for indefinite display; PEAK and TRACK modes controlled by front-
panel button, rear-panel logic input, or computer-port command
Analog Output:
±5 V, microprocessor driven and scaled to user-specified range;
±1 mV resolution
Accuracy:
0.1% of current voltage reading ±2 mV
Output Bandwidth:
1 Hz maximum*
Frequency Characteristics and Step-Response Settling Times:
See "Analog
Filtering for Displayed Reading" for the Model 3510, above
A.2.b M
ODEL
3530 LVDT C
ONDITIONER
Full-Scale Input Ranges:
"Normal": 0-150, 0-300, and 0-600 mV/V; "Long-stroke":
0-1, 0-2, and 0-4 V/V
Excitation:
Nominal 3 V-AC (rms) at 3280 Hz; 40 mA max
Common-Mode Range:
±5 V operating, ±12 V without instrument damage
Common-Mode Rejection Ratio:
Infinite at DC and 60 Hz; -60 dB at 3 kHz
Input Impedance:
Normal-Mode: 400 k
Ω
; Common-Mode: 100 k
Ω
Offset:
Initial: ±0.25% of full scale; vs. Temperature: ±20 ppm/°C; vs. Time: ±10
ppm/month
Gain Accuracy:
±0.02% of full scale ± 1 count LSD, typical, following calibration
Gain Stability:
Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading:
5-pole filter with selectable low-pass
corner frequency of 5, 10, or 20 Hz
Corner Frequency
Response at . . .
5 Hz
10 Hz
20 Hz
- 3 dB
5 Hz
10 Hz
20 Hz
-60 dB
32 Hz
65 Hz
125 Hz
Step Response Settling Time
(Full-Scale Output)
5 Hz
10 Hz
20 Hz
To 1% of final value
250 msec
125 msec
60 msec
To 0.1% of final value
400 msec
200 msec
100 msec
To 0.02% of final value
600 msec
300 msec
170 msec
A.3
3500 S
ERIES
S
PECIFICATIONS
A
A.2 I
NDIVIDUAL
C
ONDITIONER
S
PECIFICATIONS
: M
ODEL
3530
* Due to analog input bandwidth, settling times are input-limited also.
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...