Corner Frequency
Response at . . .
2.5 Hz
5 Hz
10 Hz
- 3 dB
2.5 Hz
5 Hz
10 Hz
-60 dB
15 Hz
32 Hz
65 Hz
Step Response Settling Time
(Full-Scale Output)
2.5 Hz
5 Hz
10 Hz
To 1% of final value
450 msec
275 msec
150 msec
To 0.1% of final value
850 msec
425 msec
225 msec
To 0.02% of final value
1.3 sec
650 msec
350 msec
Digital Peak Capture:
Displays the most positive measurement reading, which is
digitally held for indefinite display; PEAK and TRACK modes controlled by front-
panel button, rear-panel logic input, or computer-port command
Internal 15-Segment Linearization
, programmable from front panel
Analog Output:
Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input
Frequency Characteristics and Step-Response Settling Times:
See "Analog
Filtering for Displayed Reading" for the Model 3540, above
Configuration, Full-Scale Range, Offset Range Adjustment, Allowable
Loading, Offset Accuracy, Span Accuracy, Offset and Span Drift:
See the
Model 3530 LVDT Conditioner, above
A.2.d M
ODEL
3560 V
OLTAGE
C
ONDITIONER
Full-Scale Input Ranges:
±0.5, ±1.0, ±2.0, ±5.0, ±10.0, and ±20.0 V-DC
Excitation:
±12 V-DC ± 5%, ±40 mA maximum
Common-Mode Range:
±50 V operating, ±100 V without instrument damage
Common-Mode Rejection Ratio:
-70 dB at DC; -90 dB at 60 Hz and 3 kHz; -80 dB
a 1 kHz
Input Impedance:
Differential: 200 k
Ω
; Common-Mode: 50 k
Ω
Offset:
Initial: ±0.02% of full scale; vs. Temperature: ±25 ppm/°C; vs. Time: ±10
ppm/month
Gain Accuracy*:
±0.02% of full scale ± 1 count LSD, typical, following calibration
Gain Stability:
Vs. Temperature: ±50 ppm/°C; vs. Time: ±20 ppm/month
Analog Filtering for Displayed Reading:
See the Model 3530 LVDT Conditioner,
above
Analog Peak Capture:
Positive analog peak with minimum full-scale input pulse
duration of 6 msec (to 1% of full-scale accuracy), 12 msec (to 0.1% of full-scale
accuracy), and 20 msec (to 0.02% of full-scale accuracy); digitally held for indefi-
nite display; PEAK and TRACK modes controlled by front-panel button, rear-panel
logic input, or computer-port command
Internal 15-Segment Linearization
, programmable from front panel
Analog Output:
Range scalable in 0.1% increments between 74.5% and 125.5% of
nominal input. May be sourced by conditioned input (A) after the FIXED ANALOG
FILTER; (B) after PEAK CAPTURE; or (C) after the SELECTABLE ANALOG FILTER
(see Fig. 15, Section 2.d)
A.5
3500 S
ERIES
S
PECIFICATIONS
A
A.2 I
NDIVIDUAL
C
ONDITIONER
S
PECIFICATIONS
: M
ODEL
3560
* Initial (uncalibrated) inaccuracy may be as great as ±0.04% of full scale. Maximum error that
could occur upon replacement of the 3560 not followed by calibration is ±0.08% of full scale.
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...