4.b
P
HASE AND
S
YMMETRY
A
DJUSTMENT OF THE
M
ODEL
3578 AC S
TRAIN
G
AGE
C
ONDITIONER
PLEASE NOTE:
THIS PROCEDURE APPLIES ONLY TO THE
MODEL 3578 AC
STRAIN GAGE CONDITIONER
. FOR ALL OTHER CONDITIONERS, YOU MAY
IGNORE THIS MANUAL SECTION.
Before you calibrate your Model 3578 for the first time, you should perform the follow-
ing PHASE AND SYMMETRY ADJUSTMENT procedure. When using the 3578 with a
Lebow 1800 Series Transducer
, you should follow the special procedure outlined at
the end of this section.
ONCE SET FOR YOUR TRANSDUCER, THIS ADJUSTMENT NEED NOT BE REPEATED
UNLESS A SIGNIFICANT CHANGE IN CABLE LENGTH OR CAPACITANCE IS
REQUIRED.
1.
Make sure your transducer connection is good (see Section 2.a.7). With the 3578
in "run" mode, load the transducer in the positive direction with a convenient
"deadweight" value which is greater than one-half of full scale.
2.
Locate the PHASE ADJUSTMENT CONTROL on the rear of the 3578 (see Fig. 20).
Adjust the PHASE control until a maximum reading is obtained on the display.
3.
Remove the transducer load.
4.
Enter SETUP MODE as explained in Section 3.a.1 and press the
button. If the unit does not now display
continue to press the
CAL
button until it does.
5.
Press the
ENTER
button. The unit will display "OFSET" (for "ZERO OFFSET"):
4.4
4
S
ETUP
: I
NSTRUMENT
C
ALIBRATION
4.b P
HASE AND
S
YMMETRY
A
DJUSTMENT OF THE
M
ODEL
3578
+
+
Symmetry
Adjustment
Control
Phase
Adjustment
Control
Fig. 20 Symmetry and Phase Adjustment Controls for the Model 3578
CAL
Summary of Contents for 3500 Series
Page 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Page 38: ...2 20 This page intentionally blank...
Page 84: ...4 22 This page intentionally blank...
Page 92: ...5 8 This page intentionally blank...
Page 112: ...B 12 This page intentionally blank...
Page 114: ...C 2 This page intentionally blank...
Page 121: ...F 3 This page intentionally blank...