Chapter 3 Repairs
Note:To check the cooling circuit, the Mixolab preferably needs to
be connected to the water via a cryostat (see opposite)
1.
Access the Mixolab's maintenance menu.
2.
Click the solenoid valve icon to feed water into the cooling cir-
cuit (
A
).
The solenoid valve icon turns green and a blue bar represent-
ing the water supply appears (
B
).
3.
Run the cooling circuit for 3 minutes; this will quickly reveal
any leaks.
Watch the instrument carefully, a leak may appear (
c
), IN
WHICH CASE click the solenoid valve to stop it immediately.
4.
If a leak appears, switch off the instrument and check the
whole cooling circuit.
12: Adjusting and control operations for the Mixolab
Operation 12-6 Checking the cooling circuit
Page: 1 / 1
Consumables:
Tools:
-
-
IN
IN
OUT
OUT
Cryostat
A
B
C
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...
Page 134: ...134 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...