![Chopin Mixolab 2 Maintenance Manual Download Page 63](http://html1.mh-extra.com/html/chopin/mixolab-2/mixolab-2_maintenance-manual_2602981063.webp)
63
Maintenance manual
09/2016
Mixolab 2
Chapter 3 Repairs
8.
Thread the new detector's cable through a new coupling
washer.
9.
Insert the cable via the base of the tank.
10. Screw the detector bracket on from the inside using the
MIX-300/OUT tool.
11.
Fit the floater and lock it in place with the triangle retaining ring
(
B
).
12. Connect the detector's connector (
A
).
13. Calibrate the temperature sensor
(see op. 12-3)
.
4: Hydraulic components
Operation 4-1 Removing/replacing the double-level detector with temperature sensor
Page: 2 / 2
Consumables:
Tools:
-
-
Summary of Contents for Mixolab 2
Page 2: ......
Page 4: ......
Page 6: ...Mixolab 2 6 Maintenance manual 09 2016...
Page 8: ...8 Maintenance manual 09 2016 Mixolab 2...
Page 15: ...15 Maintenance manual 09 2016 Mixolab 2 Chapter 2 TROUBLE SHOOTING...
Page 16: ...16 Maintenance manual 09 2016 Mixolab 2...
Page 30: ...30 Maintenance manual 09 2016 Mixolab 2 Chapter 2 Trouble shooting...
Page 31: ...31 Maintenance manual 09 2016 Mixolab 2 Chapter 3 REPAIRS...
Page 32: ...32 Maintenance manual 09 2016 Mixolab 2...
Page 38: ...38 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 39: ...39 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs 2 Instructions See following pages...
Page 112: ...112 Maintenance manual 09 2016 Mixolab 2 Chapter 3 Repairs...
Page 113: ...113 Maintenance manual 09 2016 Mixolab 2 Chapter 4 SPARE PARTS...
Page 114: ...114 Maintenance manual 09 2016 Mixolab 2...
Page 116: ...116 Chapter 4 Spare parts Maintenance manual 09 2016 Mixolab 2...